Section 4
OM6530-C1-00
19 November, 2019
4-15
<Ok>
selects the highlighted profile.
<Delete>
deletes the currently selected profile. This option is only available on
user profiles and will not be present when LOCAL, GPIB, or RS232 is
highlighted.
<Save>
saves the current profile information in the selected profile.
The saved parameters are: Delay Time, Soak Time, Sensor Time, Trigger Source,
Display Resolution, Display Brightness, Auto Reverse Mode, Manual Mode,
GPIB status, RS232 status, Capacitor, Threshold, Polarity, Max Voltage, Output
Test Voltage.
<Create>
allows the creation of a new profile using the current profile
information.
<Ok>
creates a new profile with the name as entered in the text. The current
profile information will be saved in this profile.
<Scale>
allows the insertion of special text for defining the resistor value. It
toggles between M Ohms, G Ohms, T Ohms and P Ohms.