Section 5
OM6530-C1-00
19 November, 2019
5-34
5.6.109.TRACe:TSTamp:TYPE
RELative | RTClock
This allows the time of each entry in the trace buffer to be displayed either as the date and
time or as relative time from the last
<TRACe:CLEar>
command or the last detail clear
on the instrument.
5.6.110.TRACe:TSTamp:TYPE?
Typical response:
RTClock
5.6.111.TRIGger:SOURce
MANual | BUS | EXTernal | CONTinuous
<BUS>
is used if individual measurements wish to be controlled with the
<*TRG>
remote command.
5.6.112.TRIGger:SOURce?
Typical response:
Continuous
5.6.113.TRIGger:DELay
<value>
; 0-999999.999 seconds
This allows the instrument to provide any internal delay between each measurement. It is
normally set to zero (0) but is used to create a delay between measurements especially if
they are very fast measurements.
5.6.114.TRIGger:DELay?
Typical response:
0.0
5.6.115.TRIGger:SOAK
<value>
; 0-999999.999 seconds
This allows the instrument to allow additional settling time between reversals of polarity.