Section 4
OM6530-C1-00
19 November, 2019
4-36
actual variance (ppm) and standard deviation (ppm) is displayed. If for some
reason the routine continues until the maximum retries has been reached, then the
last window of samples is used to generate the actual variance (ppm) and standard
deviation (ppm) that is displayed. The values are not stored in the 6530 non-
volatile memory. When the calibration is complete it will allow the user to
<save>
the values or to select another capacitor and
<Calc>
the calibration
coefficient.
4.9.2.2.Calibration Coefficient Values
This section allows the editing of the calibration coefficients values.
4.9.2.2.1.Capacitor Coefficients
Pressing the
<Capacitors>
key will display the coefficients stored for each
capacitor. By pressing the
<Edit>
key the numbers may be altered by pressing a
new sequence of numbers followed by pressing the
<Ok>
key. Pressing the
<PREVIOUS>
key exits to the SOFCAL Calibrate Menu.
4.9.2.2.2.Threshold Coefficients