Section 4
OM6530-C1-00
19 November, 2019
4-31
4.9.2.
Calibrate
<Utilities>
is used to calibrate the capacitors, thresholds, protection resistor and
shunt.
<Cal. Vals>
is used to enter or modify the test volts, threshold, protection resistor
and shunt correction values.
<Ref. Vals>
is used to enter/modify the instrument serial number, calibration
date, auto cal parameters, zero parameters and shunt parameters.
4.9.2.1.Utilities
It should be noted that the instrument may not reach a stable state if it is running
in a environment which has climate fluctuations or high levels of electrical or
other interference. In some cases it may be necessary to change the variance or
the standard deviation from the system default values as displayed in the “Ref-
Vals” submenu. If any of the calibration routines is unable to meet the
specification then an error message <
*** Sample Maximum Exceeded>
will be
displayed and the procedure will be aborted.