Baseline Configuration
The baseline configuration button may be used to select the point at which ear
canal volume (Vea) is calculated. When the button is selected, a drop-down list
is presented. When set to Auto, Vea for a 226 Hz probe tone tympanogram is
calculated at the positive tail and for a 1000 Hz probe tone tympanogram, Vea is
calculated from the negative tail. When Positive Tail is selected, Vea for all
probe tones is calculated at the positive tail. When Negative Tail is selected, Vea
for all probe tones is calculated at the negative tail.
Peak Indicator
The Peak Indicator Display Options button determines if the tympanogram peak
is marked. When the button is checked, a vertical line is drawn through the peak
of the tympanogram to indicate the peak and corresponds to the point whose
values for peak is displayed in the results.
Comments
The pencil icon opens a comments window (must use external keyboard to utilize
comment section). Comments may be entered from any test screen and it is
possible to review and edit comments from any test screen.