![GSi TYMPSTAR PRO User Manual Download Page 124](http://html1.mh-extra.com/html/gsi/tympstar-pro/tympstar-pro_user-manual_2275120124.webp)
Rate of Pressure Change
12.5, 50, 200, 600 daPa/sec +/- 10%
Acoustic Reflex Activating Stimulus System
Pure Tone Stimulus Signals
Frequency (Hz)
Min Level
(dB HL)
Max
Level
(dB HL)
Ipsi
Max Level
(dB HL)
Contra-
Insert
Max Level
(dB HL)
Contra-
Headset
Frequency
Accuracy
250
35
85
105
105
+/- 3%
500
35
100
110
120
+/- 3%
1000
35
105
120
120
+/- 3%
2000
35
105
120
120
+/- 3%
4000
35
100
115
120
+/- 3%
Broad Band
Noise
35
93
108
115
Low Band
Noise
35
94
109
115
High Band
Noise
35
90
110
115
Frequency (Hz)
RETSPL
IPSI Probe
RETSPL
Contra Insert
RETSPL
Contra Headset
250
24.5
14.0
27.0
500
9.5
5.5
13.0
1000
6.5
0.0
6.0
2000
12
3.0
8.0
4000
3.5
5.5
9.0
Broad Band Noise
7.0
7.0
7.0
Low Band Noise
9.0
9.0
9.0
High Band Noise
7.0
7.0
7.0
Zero level reference values (250 Hz-4000 Hz) are a GSI Standard and come from
an internal study performed according to ISO 389-9. BBN, LBN and HBN
reference values were determined by GSI using the “Threshold Determination
Method”.
Pure Tone Harmonic Distortion
< 5% at 500 Hz to 2000 Hz, 100 dB HL
< 5% at 4000 Hz, 80 dB HL
< 10 % at maximum dB HL settings
NOTE:
Levels are reduced as a function of volume at a rate of 1 dB SPL for
each .1 ml. Level reduction begins at 1.2 ml.