Reflex Test
Pressing the Reflex button displays the reflex screen. Pressing the Reflex button
a second time displays the Reflex Decay screen. The reflex button acts as a
toggle button switching between the reflex screen and the reflex decay screen.
The reflex screen is show below.
The reflex screen presents information for conducting diagnostic acoustic
stapedial reflex tests. The reflex collection data are displayed as graphs with
different probe tones stored as tabs on the left side of the reflex graph. The tabs
at the top of the graph are used to select the ear to be tested. The area to the right
of the graphic area displays the admittance and pressure meters along with the
eliciting stimulus information.
Reflex test data is collected in the scratch pad (upper left graph area) and is not
part of the session until it is stored in the Reflex Results area. There are several
ways data may be stored in the Reflex Results area. The data is automatically
stored at the end of a Threshold Seek (TS) test. In this case, the “TS” indicator
will be labeled on the graph and a threshold that meets the criteria will be
automatically selected by the system. The threshold is indicated by a box/line