tympanogram. The Vea column displays the ear canal volume (in cm
3
). The
Peak column displays the pressure (daPa) and admittance (mmho) value of the
peak of the tympanogram. The Tymp Measure column will display either the
gradient in mmhos or the tymp width in daPa depending on the settings in the
configuration. The Type column displays tympanogram classification type that
the user selects from the Set Type link in the column.
Set Type
The Set Type option displays a dialog with the tympanogram types that are
available to be selected. The choices for the tympanogram type are defined in
the PC configuration application (TympStar Pro Config App).
Reflex Parameter Buttons
Up to four reflexes may be recorded on the Screener display and there are two
parameter buttons for each reflex. For each reflex the choice for the stimulus ear
and stimulus frequency may be selected. The Stim. Ear button selects which ear
the reflex activation stimulus is presented. This button presents a selection of the
options; Ipsi, Contra and None. When Ipsi is selected, the reflex stimulus is
presented by the probe and therefore to the same ear as the probe. When Contra
is selected, the reflex stimulus is presented to the contra probe transducer which
is placed in the non-probe ear. The Frequency button presents a selection of the
reflex activation stimulus frequency (Hz) options of 500, 1000, 2000, and 4000
Hz and BBN (broad band noise).