sweep goes from negative pressure to positive pressure). The peak of the
tympanogram may be indicated by the arrow and line if the Peak Indicator is
selected in Settings. The peak may be manually set by the user using the Zoom
function.
Probe Status Indicator
The probe status indicator indicates the state of the probe and displays messages
regarding the current test. On the left side of the indicator, the two circles will be
colored to match the LED lights on the probe box. The probe may be in the
following states:
Indicator Lights
Status
Blinking Green
Ready
Solid Green
Test in Progress
Solid Green
Hold – ready to continue test
Solid Amber
Occlusion
Blinking Amber
Leak (or possible internal error on the device)
Admittance and Pressure Meters
The admittance meter displays the real time admittance as it is being recorded
from the probe in the ear canal. The units are millimhos (mmho). The current
value is displayed at the top left of the meter.
The pressure meter displays the real time pressure as it is being recorded from the
probe in the ear canal. The units are decapascals (daPa). The current value is
displayed at the top left of the meter. The pressure range is controlled from the
pressure range parameter for the test.
Manual
Pressing the Manual button initiates a tympanogram collection where the
pressure knob is used to manually increase or decrease the pressure. When
performing manual measurements, multiple tympanograms are displayed until
the Stop key is pressed. The graphs from the last three measurements are saved
and presented in the table of results.