background image

English (GB)

22

9.7.1 Result file

The result file is in .txt format and structured as follows:

10. Maintenance

10.1 Cleaning

The following components of the DID require periodical checks 
and cleaning, the interval depends on the quality of the water in 
which they are used. We suggest to check these components on 
a monthly basis.

10.1.1 Inlet strainer

The strainer includes a filter that collects coarse material and can 
foul. Therefore, a periodical check of the strainer is necessary. 
We recommend to clean the filter in the following intervals:
• monthly
• if the flow drops, although all valves are open, the pump (if 

installed) is working properly and the inlet pressure is suitable.

Cleaning the inlet strainer

1. Turn off the water flow.
2. Unscrew the cap from the strainer.
3. Remove the filter.
4. Rinse the filter with drinking water.
5. Remove any coarse materials from the strainer housing.
6. Reinsert the filter.
7. Screw the cap back onto the strainer.

10.1.2 Flow detector

The flow detector contains inner parts that can be affected by 
fouling. It case the flow detector does not respond to flow 
changes, it can be dismantled from the DID and carefully cleaned 
with water and a soft brush.

Dismantling the flow detector

1. Turn off the water flow.
2. Disconnect the CU 382 control unit from the power supply.
3. Unscrew the hose connection.
4. Remove the hose.
5. Disconnect the wires of the flow detector from the CU 382.
6. Unscrew the flow detector.
7. Clean the flow detector carefully with water and a soft brush.
8. Screw in the flow detector back.
9. Reconnect the wires of the flow detector to the CU 382.
10. Refit the hose.
11. Screw on the hose connection.

10.1.3 Sensor

See the manual "Sensors V1" supplied with the sensor.

10.1.4 Control unit CU 382

If necessary, clean the surfaces of the CU 382 control unit with a 
damp and clean cloth.

Timestamp

Seconds since 1970/1/1

Time

YearMonthDay-HourMinuteSecond

P

Marks the beginning of the parameter data 
block

Parameter value

Value of the parameter

Parameter status

Status of the parameter

Parameter private 
status

Private status of the parameter

Upper alarm 
triggered

If upper alarm was triggered = "U" else "-"

Lower alarm 
triggered

If lower alarm was triggered = "L" else "-"

C

Marks the beginning of the controller data 
block

Controller output 
[%]

Output of the controller

Disturbance input

Input of disturbance control

Controller status

Status of the controller
0x0001 - General error
0x0002 - Input error
0x0004 - Output error
0x0008 - Disturbance error

Controller 
operating mode

Operating mode of the controller
0 - "

Stopped

"

1 - "

Running

"

2 - "

Hold

"

WARNING
Electric shock

- Switch off the power supply before connecting the 

power supply cable and relay contacts.

- Do not dismantle the control unit.
- Maintenance and repair must only be carried out 

by qualified persons.

- Observe the local safety regulations.

WARNING
Toxic material

- Parts of the DID can be contaminated with 

dangerous chemicals or pathogenic germs.

- If the DID has been in contact with dangerous 

chemicals or pathogenic germs, wear protective 
clothing and take all necessary precautions to 
prevent endangering your health when installing or 
dismantling the DID. 

Summary of Contents for DID BF1

Page 1: ...DID Installation and operating instructions GRUNDFOS INSTRUCTIONS Other languages http net grundfos com qr i 99037650...

Page 2: ...ety instructions 3 3 Receiving the product 3 3 1 Inspecting the product 3 3 2 Transporting the product 3 4 Installation 3 4 1 Installation location 3 4 2 Installation of the DID 3 4 3 Hydraulic connec...

Page 3: ...installation of the DID with flow cell on a flat wall See also sections 8 Product introduction and 12 4 Dimensions DID with sensor for tank immersion The CU 382 control unit of the DID with sensor for...

Page 4: ...ease of the flow 5 Starting up the product 5 1 Initial startup 1 Connect all sensors to the CU 382 control unit See also section 6 2 Connecting the sensors to the CU 382 control unit 2 Slowly open the...

Page 5: ...ialisation process Connect only the sensor you want to initialise to the CU 382 An individual address is allocated to the initialised sensor If no sensor was connected before the CU 382 jumps into the...

Page 6: ...f DID with flow cell Fig 5 Components of DID with immersed sensor TM06 8088 3416 9 5 4 3 2 1 11 10 10 12 13 14 15 15 15 15 7 6 6a 8 18 T OK W X S 19 6c 6b TM06 8089 3416 Pos Description 1 Water inlet...

Page 7: ...n marks 8 QR code 9 Type 10 Voltage V 11 Frequency Hz 12 Power consumption W 13 Humidity limits 14 Temperature limits 1 2 3 5 7 8 10 11 12 13 14 4 6 9 TM06 8610 0817 Pos Description 1 s can product na...

Page 8: ...Moves the marker to the line above The first character of the selected line is a triangle Increases values Function button Enters the setup menus Disables or deletes values Left button Moves the mark...

Page 9: ...05B1 pH 15 YYYY MM DD Serial 9999999 9 49 Type PID Running ls below Waiting 1 s Temp pH 4 20mA 2 lower alarm limit YYYY MM DD hh nn ss 22 1 C AV SP 6 86 7 00 ext General setup Parameter setup Controll...

Page 10: ...0mA digital outputs PULSE or PWM AV SP AV is the actual value input parameter reading SP is the setpoint desired value of the input parameter reading Source of the setpoint External Manual AP001 A1 1...

Page 11: ...he Up and Down buttons Confirm with OK To add a MODBUS sensor select Add MODBUS sensor with the Up and Down buttons Confirm with OK The MODBUS port for a new sensor is searched and the new sensor is a...

Page 12: ...e configured in the MODBUS slave menu Changing the Modbus address 1 Select Address with the Up and Down buttons Confirm with OK 2 Change the address value with the Up and Down buttons Confirm with OK...

Page 13: ...Digital Input Read in all digital inputs Internal Sensors Read in all internal sensors USB Measure the performance of the USB stick Date Time Set the time and the date Hardware Read out the hardware r...

Page 14: ...he second reference solution Select the value in the Lab 2 field with the Up and Down buttons Confirm with OK Use a temperature compensated value 11 Select Sample 2 with the Up and Down buttons Confir...

Page 15: ...ith the Up and Down buttons Confirm with OK 2 The default display settings are loaded from the sensor All modifications of the display settings are documented in the logbook of the CU 382 control unit...

Page 16: ...ut settings menu a parameter can be allocated to an analog output or a digital output Analog outputs Parameter values between these two limits are calculated according to a linear calibration Assignin...

Page 17: ...ect Period s with the Up and Down buttons Confirm with OK 2 Change the value for the duration of one pulse with the Up Down Left and Right buttons Confirm with OK 9 4 7 Parameter info The Parameter in...

Page 18: ...e type to Manual 1 Select Type with the Up and Down buttons Confirm with OK 2 Select Manual with the Up and Down buttons Confirm with OK Changing the setpoint of the controller 1 Select Setpoint with...

Page 19: ...1 Select D factor with the Up and Down buttons Confirm with OK 2 Change the disturbance factor with the Up Down Left and Right buttons Confirm with OK Changing the direction of the controller 1 Select...

Page 20: ...er that can be changed is the maximum of pulses min for 100 output Changing the maximum of pulses min for 100 output 1 Select Pulses Min with the Up and Down buttons Confirm with OK 2 Change the maxim...

Page 21: ...all logfiles to the USB stick 1 Select Copy logfile with the Up and Down buttons 2 Confirm with OK to copy the data to the USB stick Deleting all logfiles 1 Select Delete logs with the Up and Down but...

Page 22: ...arefully with water and a soft brush 8 Screw in the flow detector back 9 Reconnect the wires of the flow detector to the CU 382 10 Refit the hose 11 Screw on the hose connection 10 1 3 Sensor See the...

Page 23: ...layed DID All hoses and fittings are tight CU 382 control unit The digital output of the flow detector is ok Outlet Water is flowing out of the outlet hose Water is visible within the hoses Flow cell...

Page 24: ...s no flow a The inlet valve is closed Open the inlet valve b The inlet strainer is blocked Clean the inlet strainer 8 Water is not visible in the hoses a The inlet valve is closed Open the inlet valve...

Page 25: ...n data Internal thread of inlet and outlet Rp1 2 Size of inlet and outlet connection thread G 5 8 Inlet outlet hose connections metric ID OD mm 4 6 6 9 6 12 or 9 12 Inlet outlet hose connections inch...

Page 26: ...age VAC 240 Maximal switchable load A 6 Maximal number of operating cycles VA 600 Required external protection Depending on the load additional snubber circuits may be required Description Unit D 320...

Page 27: ...5 Terminal connections 4 x M12 cable gland entry one used by flow switch 1 x M16 cable gland entry IP rating IP65 Memory MB 512 industrial grade SLC Display mm 128 x 64 graphical display 70 x 40 viewi...

Page 28: ...3 FCL2 ORP 4 710 8 765 DID 3 BF3 FCL2 PH ORP 4 960 9 165 DID 1 BF1 PH 4 250 8 085 DID 1 TI1 PH 2 075 5 875 DID 1 TI1 ORP 2 075 5 875 DID 3 TI2 ORP PH 2 850 6 800 DID 3 BF3 ORP PH 4 700 8 755 DID 1 BF1...

Page 29: ...tact the nearest Grundfos company or service workshop TM06 5897 0216 TM06 6025 0416 Fig 12 Dimensions of CU 382 control unit front and side in mm Fig 13 Dimensions of CU 382 control unit for wall moun...

Page 30: ...t of the sensors The sensors must always be immersed in water Avoid siphon effect Use a vented outlet or a pressure loading valve Avoid power loss of the sensors that are equipped with electrolyte and...

Page 31: ...Documents as listed Type of product Measurement Process Control Name of product DID Model number D 320 GF1 230 D 320 GF3 230 Low Voltage Directive 2014 35 EU Standard used EN 61010 1 2010 RoHS2 2011...

Page 32: ...an GRUNDFOS Pumps K K 1 2 3 Shin Miyakoda Kita ku Hamamatsu 431 2103 Japan Phone 81 53 428 4760 Telefax 81 53 428 5005 Korea GRUNDFOS Pumps Korea Ltd 6th Floor Aju Building 679 5 Yeoksam dong Kangnam...

Page 33: ...NDFOS Pumps K K 1 2 3 Shin Miyakoda Kita ku Hamamatsu 431 2103 Japan Phone 81 53 428 4760 Telefax 81 53 428 5005 Korea GRUNDFOS Pumps Korea Ltd 6th Floor Aju Building 679 5 Yeoksam dong Kangnam ku 135...

Page 34: ...1168 The name Grundfos the Grundfos logo and be think innovate are registered trademarks owned by Grundfos Holding A S or Grundfos A S Denmark All rights reserved worldwide Copyright Grundfos Holding...

Reviews: