Operation
GENO
-Multi BS
Order no. 004 523 950-gb Edited by: KONS-rjau-mrie G:\BA-523950-GB_KD-F.DOC
-CONFIDENTIAL-
F-7
2.1.2 Programming parameters
Key functions with parameter programming – requirement: Basic
display of irradiance [W/m²].
Key Display
Press the key > 1 sec.: Access to parameter programming
(opt. temperature flushing system).
Open the parameter for reprogramming – the value begins to flash.
Save the parameter – the value stops flashing.
Reduce the opened parameter (while value flashes).
Return to the previous parameter.
Increase the opened parameter (while value flashes).
Return to the next parameter.
Close the opened parameter without saving – the old set value is main-
tained.
Jump back to the basic display of irradiance.
Open the display of the code query "Code C.000" >> access to code-
protected programming levels.
2.2 User programming level
The limit value for the optional temperature flushing system is
saved in the user programming level:
Setting range 10 … 60 °C / factory setting 38 °C
When the limit value is reached the rinsing solenoid valve opens
and the display shows Er 3, the zero-potential contact ALARM
opens.
The rinsing solenoid valve remains open until the water has
cooled down by 2 °C.
The malfunction can be manually acknowledged and rinsing is
not stopped by this.
+
+