Page
47 www.greyline.com
5.3
Warning & Status Messages
FLOW RATE ERRORS
No flow signal
Interpretation:
This message appears when the transducers cannot send or
receive signals to each other.
Response:
Firstly check that all cables are connected, transducers are on the pipe
correctly with sufficient couplant on the face.
This condition could also be due to a partially empty pipe, aerated liquid, particulate
content too high or when the condition of the pipe being measured is poor.
Flow signal is poor
Interpretation:
This warning appears when the signal is lower than 25%.
Response:
This could be due to an application problem, a poor quality pipe – see
also the conditions for No flow signal (above). Check for sufficient couplant.
Zero cut-off error!
Interpretation:
You have entered an out-of-range value in the Zero cutoff field in
the Options menu.
Response:
Enter a valid number.
Damping error!
Interpretation:
You have entered an out-of-range value in the Damping field in the
Options menu.
Response:
Enter a valid number.
Totaliser beyond maximum!
Interpretation:
The totaliser has overflowed its maximum count. The counter will
roll-over and restart from zero but this message alerts you to the fact.
Response:
Reset the totaliser as described in
Paragraph 3.6.1
.
PULSE ERRORS
Pulse Rate > Max
Interpretation:
The flow rate exceeds the capability of the pulse output – i.e. too
many pulses per second are required than can be achieved.
Response:
Narrow the pulse width time or increase the volume per pulse, as
described in
Paragraph 3.7.7
.
Pulse volume error!
Interpretation:
You have entered an out-of-range value in the Pulse volume error
field in the PULSE OUTPUT menu – see
Paragraph 3.7.7
.
Response:
Enter a valid number.
Pulse width error
Interpretation:
You have entered an out-of-range value in the Pulse width error
field in the PULSE OUTPUT menu – see
Paragraph 3.7.7
.
Response:
Enter a valid number.