Page
31 www.greyline.com
Monitoring the logged events as a graphic
REAL TIME LOGGER
DD-MM-YY HH:MM:SS
4-20mA O/P is ON.
Dim: mm
Unit
: l/min
Log name
: Quickstart
Log data to
: Memory
Logging interval
: 10 seconds
Start date & time
: dd-mm-yy hh:mm:ss
Stop date & time
:
dd-mm-yy hh:mm:ss
10. If you want to view the logging progress graphically
rather than as text, press the Logger function key
to get to the REAL TIME LOGGER screen.
11. Select Graph Y axis max and enter a value equal
to the expected maximum flow.
12. Select View log as graph. This will open the VIEW
LOG AS GRAPH screen shown below.
13. You can select the data at any point in time by
scrolling along the time axis of the graph using the
LEFT/RIGHT scroll keys (5 & 6). The data values
for the selected point are then shown to the left of
the graph. Keep the scroll key pressed for the
cursor to move automatically.
MM DD HH MM SS
Remaining time
: mm dd hh mm ss
Memory Rollover
: Overwrite
Graph Y axis max.
: 50
View log as text
View log as graph
START NOW
Set Auto start
Clear log
Exit
VIEW LOG AS GRAPH
DD-MM-YY HH:MM:SS
Dim: mm
50.00
Log: Quickstart
Point: XXX/XXX
Value: xx.xx
dd-mm-yy hh:mm:ss
–50.00
14. If you want to change the value of the Y axis to
improve the resolution of the graph press the
UP/DOWN scroll keys. The maximum values will
then increase/decrease as appropriate.
15. To exit the VIEW LOG AS GRAPH screen and
return to the REAL TIME LOGGER screen press
ENTER.
Monitoring the logged events at a later time
16. If you wish to monitor the logging progress at any time while you are operating in FLOW READING mode,
press the Logger function key as described above to switch to the REAL TIME LOGGER screen. You can
then select View log as text or View log as graph, as required, without interfering with the logging operation.
Stop logging
REAL TIME LOGGER
DD-MM-YY HH:MM:SS
4-20mA O/P is ON.
Dim: mm
Unit
:
l/min
Log name
:
Quickstart
Log data to
: Memory
Logging interval
:
10 seconds
Start date & time
:
dd-mm-yy hh:mm:ss
Stop date & time
:
dd-mm-yy hh:mm:ss
17. From the FLOW READING screen, press the
Logger function key to access the REAL TIME
LOGGER screen.
18. Select STOP NOW to cease logging.
Note that the START NOW and STOP NOW
options toggle depending on which was last
selected.
19. Select Exit to return to the FLOW READING
screen.
Note that the logged data will remain stored in the
instrument’s memory and can be accessed at any time
as described above.
MM DD HH MM SS
Remaining time
: mm dd hh mm ss
Memory Rollover
: Overwrite
Graph Y axis max.
: 50
View log as text
View log as graph
STOP NOW
Set Auto start
Clear log
Exit