3.6.2 How to set up the basic logging application to memory
This procedure shows you how to set up a basic logging session under manual start/stop control. The logged
data is saved to the instrument’s memory and can be downloaded to a PC at a later time.
Starting point
This procedure assumes that the Portaflow has been correctly installed and is operating in the FLOW READING
mode.
Setting up and starting the logging
1. With the Portaflow now operating in FLOW READING mode, check that the indicated flow units are the same
as those you want to appear on the logger output (e.g. l/min).
REAL TIME LOGGER
DD-MM-YY HH:MM:SS
4-20mA O/P is ON.
Dim: mm
Unit
:
l/min
Log name
: Quickstart
Log data to
: Memory
Logging interval
: 10 seconds
Start date & time
: dd-mm-yy hh:mm:ss
Stop date & time
: dd-mm-yy hh:mm:ss
2. Press the Logger function key to access the REAL
TIME LOGGER screen.
3. Select Log data to and choose Memory to store the
log to the internal memory.
4. Select Logging interval and enter the required
period.
5. Select Clear log if you want to discard any existing
logging data.
6. Select START NOW. Logging should begin, and a
Logging to memory message should appear on the
second line of the display.
7. Select View log as text to check that logging is
taking place.
MM DD HH MM SS
Remaining time
: mm dd hh mm ss
Memory Rollover
: Overwrite
Graph Y axis max.
: 50
View log as text
View log as graph
START NOW
Set Auto start
Clear log
Exit
VIEW LOG AS TEXT
DD-MM-YY HH:MM:SS
Logging to memory
Dim: mm
Log:Quickstart
Date
Time Flow
DD-MM-YY hh:mm:ss xxx.xx l/m
DD-MM-YY hh:mm:ss xxx.xx l/m
DD-MM-YY hh:mm:ss xxx.xx l/m
DD-MM-YY hh:mm:ss xxx.xx l/m
8. The VIEW LOG AS TEXT screen displays the log
events in date-stamped chronological order.
Note:
Press key 5 to go to the top line of logged
data or key 6 to go to the bottom line.
9. Press the ENTER key to return to the REAL TIME
LOGGER screen. Then select Exit to return to the
FLOW READING screen. Logging will continue to
take place in the background.
--------
DD-MM-YY hh:mm:ss xxx.xx l/m
Key Point:
There can be only one set of logged data per site. If a new data log is
started on a site it will clear the existing data.