16. Check that the indicated signal strength on the left
of the screen is at least 2 bars (ideally 3 or 4). If
less than 2 bars are shown it indicates there could
be a problem with the transducer spacing,
alignment or connections; or it could be due to an
application problem.
FLOW READING
DD-MM-YY HH:MM:S
Signal
Dim: mm
0.000
l/min
+Total: 0.00
litres
–Total: 0.00
litres
Flow monitoring
The FLOW READING screen is the one most used during normal monitoring operation. It shows the
instantaneous fluid flow together with totalised values (when enabled). In this mode you can select the flow rate
measurement units by pressing keys 7 (litres), 8 (Gallons) or 9 (m³), or change the display to show velocity by
pressing key 4.
3.4
Managing Named Sites
If you want to monitor a particular site location frequently you can set up a named ‘Site’ to store the installation
details, such as pipe dimensions and material, required to set-up the Portaflow PT500 system. These can then be
recalled later when revisiting that particular location.
The instrument can store up to 20 sites, the first site is reserved for QUICK START and cannot be renamed;
subsequent sites are initially named EmptySite1 through to EmptySite19.
3.4.1 Setting up a new site
1. Select View Edit Site Data from the MAIN MENU.
2. Select Choose from list of sites.
3. Select one of the EmptySites from the presented
list (e.g. EmptySite 1 as shown).
4. Select Site name and press ENTER.
VIEW EDIT SITE DATA
DD-MM-YY HH:MM:SS
Dim:
mm
Choose from list of sites
Site name
: EmptySite1
Dimension units
: mm
Pipe outside diameter
: 58.00
Pipe wall thickness
: 4.00
Pipe lining thickness
: 0.00
5. This opens the ALTER NAME screen.
6. Select Alter the Site Name and you will be
presented with a screen which allows you to enter
a new name in much the same way as when
composing a mobile text message.
7. On completion press ENTER then select Exit.
This will take you back to the VIEW EDIT SITE
DATA screen.
ALTER NAME
DD-MM-YY HH:MM:SS
Site name
: EmptySite1
Do you wish to:–
Alter the Site Name
Exit