2  Designated Use 

The GMH 3651 is measuring dissolved oxygen in Water.  
The measuring is performed by means of a suitable oxygen sensor (included in supply in standard 
instrument) connected to the Mini-DIN socket, the sensor measures at the membrane at the tip of the sensor. 
Due to the properties of the sensor, it has to be calibrated regularly (e.g. at fresh air = 20.95%) to get precise 
values. If the sensor is used up, this will be detected during the calibration, the sensor has to be regenerated 
or replaced before continuing with measuring. 

The safety requirements (see below) have to be observed.  
The device must be used only according to its intended purpose and under suitable conditions.  
Use the device carefully and according to its technical data (do not throw it, strike it, …) 
Protect the device from dirt. 

3  Safety 

3.1  Safety signs and symbols 

Warnings are labeled in this document with the followings signs: 

 

DANGER 

        

Caution!

 This symbol warns of imminent danger, death, serious injuries 

and significant damage to property at non-observance. 

 

Attention!

 This symbol warns of possible dangers or dangerous 

situations which can provoke damage to the device or environment at 
non-observance. 

 

Note!

 This symbol point out processes which can indirectly influence 

operation or provoke unforeseen reactions at non-observance. 

3.2  Safety guidelines 

This device has been designed and tested in accordance with the safety regulations for electronic devices. 
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures 
and special safety advises given in this manual will be adhered to when using the device. 

1.  Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected 

to any other climatic conditions than those stated under "Specification". 
If the device is transported from a cold to a warm environment condensation may cause in a failure. In 
such case make sure the device temperature has adjusted to the ambient before trying a new start-up. 

2.    

 

If there is a risk whatsoever involved in running it, the device has to be switched off 
immediately and to be marked accordingly to avoid re-starting. 
Operator safety may be a risk if: 
- there is visible damage to the device 
- the device is not working as specified 
- the device has been stored under unsuitable conditions for a longer time. 
In case of doubt, please return device to manufacturer for repair or maintenance. 

3. 

 

 

Do not use these products as safety or emergency stop devices or in any other application 
where failure of the product could result in personal injury or material damage. 
Failure to comply with these instructions could result in death or serious injury and material 
damage. 

 

4. 

 

 

 

This device must not be used at potentially explosive areas! The usage of this device at 

potentially explosive areas increases danger of deflagration, explosion or fire due to 
sparking.  

 

   

 DANGER

 

 

       

 

 

DANGER

 

 
 
 

       

  

 

DANGER

 

 
 
 

Summary of Contents for GMH 3651

Page 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 0 Operating Manual GMH 3651 Oxygen meter for dissolved oxygen with integrated temperature and pressure...

Page 2: ...IBRATION AL 1 PT 11 8 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 12 8 3 EVALUATION OF SENSOR STATE ELE 12 9 OPERATION OF THE LOGGER 13 9 1 FUNC STOR STORING SINGLE MEASUREMENT SETS 13 9 2 FUNC CYCL AUTOM...

Page 3: ...s device has been designed and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safet...

Page 4: ...s to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V power supplies often hav...

Page 5: ...4 ok arrow Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 SAL arrow Signals that the salinity correction is aktivated 7 Logg arrow Sh...

Page 6: ...lip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attached to a belt etc The device with pop up clip at position 90 can b...

Page 7: ...h in meter only with depth accessory TMV 3600 DEP Depth measuring activated Second display always shows depth in meter only with depth accessory TMV 3600 U N T T C All temperatures in degree Celsius e...

Page 8: ...rement 0 0 SET AL Set Alarm AL 1 ON NO SO Monitoring Oxygen Alarm on with buzzer Alarm on without buzzer OFF No alarm monitoring for oxygen AL N ONC Monitoring Oxygen Concentration in P 02 Monitoring...

Page 9: ...to ageing Therefore we recommend to maintain the electrode at monthly intervals p r t Electrode maintenance Please note Make it a rule to always store the electrode in a humid environment in the stora...

Page 10: ...xygen concentration requires entering the salinity of the medium p r t Configuration The correction of salinity is based on media on a water basis whose chemical content is similar to sea water The co...

Page 11: ...ctrode is exposed to air with a relative atmospheric humidity of 100 Proceed as follows Put some distilled water in a bottle Close bottle and generate a water steam saturation 100 rel atmospheric humi...

Page 12: ...blinks no valid reference is recognised by the instrument As soon the values for oxygen and temperature are stable the calibration of the second point will be finished At 2 point calibration the calib...

Page 13: ...f data sets 1000 A data set contains oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute p...

Page 14: ...The recording automatically will be stopped Stopping the recording manually By pressing Store key 6 the recording can be stopped manually Then the following choice appears The selection can be made by...

Page 15: ...nels oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute pressure in hPa abs or mmHg abs o...

Page 16: ...rtly switch between ELE total electrode state evaluation in SL 1 Slope at 0 Air saturated water 1 SL 2 Slope at air saturated water 100 1 Date time display of data set or Change between the different...

Page 17: ...or Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery will be exhausted soon Further measurements a...

Page 18: ...000 data sets with measuring point input Alarm Buzzer visual interface 2 channels selectable oxygen unit and temperature Additional functions Min max hold auto hold Housing Break proof ABS housing Pro...

Reviews: