H74.0.05.6C-06
Operating Manual GMH 3651
page 7 of 18
5 Configuration
Some menu points depend on current device settings.
To change device settings, press
„Menu“
for 2 seconds. This will activate the configuration menu (main
display: “Set”). Pressing
„Menu“
changes between the menus points, pressing
jumps to the
referring parameters, which can be selected with key
.
The parameters can be changed with
or
. Pressing
„Menu“
again jumps back to the main
configuration menu and saves the settings. "Quit"
finishes the configuration and returns to standard
measuring operation.
Menu
Parameter Value
Description
key Menu
key
key
or
p.r.t
SET
(ONF
Set Configuration: General configurations
U N,T
(ONC
(ONC
ppm
Display unit of O
2-
concentration ppm (identical values like mg/l)
*
(ONC
mg/l
Display unit of O
2-
concentration mg/l (identical values like ppm)
[H 2
SAT
%O
2
Main display shows O
2-
concentration (mg/l or ppm) or O
2
-Saturation in
%
(ex works setting)
P 02
hPa
Main display shows O
2-
concentration or O
2
-partial pressure in hPa
P 02
mmHg
Main display shows O
2-
concentration or O
2
- partial pressure in mmHg
RES
K ,
Best O
2
resolutions
LO
Low O
2
resolution, calm value display (standard)
SAL
OFF
Salinity correction deactivated
(ex works setting)
0.1
…
70.0
Salinity correction activated, Unit
‰ = PSU
LCD.2
T
Second. display always temperature
P
Second. display always absolute pressure
P T
Second. display alternates between temperature and abs. pressure
DEP. T
Depth measuring activated, Second. display alternates between
temperature and depth in meter (only with depth accessory TMV 3600)
DEP
Depth measuring activated, Second. display always shows depth in
meter (only with depth accessory TMV 3600)
U N,T
T
°C
All temperatures in degree Celsius (ex works setting)
°F
All temperatures in degree Fahrenheit
[AL
A,R
The sensor will be calibrated at air or gas
(ex works setting)
A Q UA
The sensor will be calibrated in water
[AL.P
1-PT
Simple one point calibration at air (
A,R
) o rat air saturated water (
A Q UA
)
2-PT / 3-PT
2 or 3-point calibration at air (
A,R
)/air saturated water (
A Q UA
), or in
oxygen/oxygen saturated water or in nitrogen/zero solution
[. INT
1
…
365
Calibration reminder period (in days)
OFF
No calibration reminder
A VTO
HLD
ON
Auto measuring value identification Auto Hold (when logger = off)
OFF
Standard hold function on key press (when logger = off)
P.OFF
1
…
120
Power-off delay in minutes. Device will be automatically switched off as
soon as this time has elapsed if no key is pressed/no interface
communication takes place. (ex works setting 20min)
OFF
Automatic power-off function deactivated (continuous operation)
0V T
OFF
Interface off -> minimal power consumption
SER
Serial interface activated (ex works setting)
DA[
Analogue output activated
ADR
0 1
,
1 1
,
2 1
,
…
9 1
Base address for serial interface communication
(ex works setting 01)
1
2
3
4
5
6
ON
OFF
max
Store
CAL
min
Set
Menu
Quit
1
2
3
4
5
6
ON
OFF
max
Store
CAL
min
Set
Menu
Quit
1
2
3
4
5
6
ON
OFF
max
Store
CAL
min
Set
Menu
Quit
1
2
3
4
5
6
ON
OFF
max
Store
CAL
min
Set
Menu
Quit
1
2
3
4
5
6
ON
OFF
max
Store
CAL
min
Set
Menu
Quit
1
2
3
4
5
6
ON
OFF
max
Store
CAL
min
Set
Menu
Quit
1
2
3
4
5
6
ON
OFF
max
Store
CAL
min
Set
Menu
Quit
1
2
3
4
5
6
ON
OFF
max
Store
CAL
min
Set
Menu
Quit