background image

H74.0.05.6C-06 

Operating Manual GMH 3651 

page 9 of 18 

6  The Oxygen Sensor 

6.1  Design of Sensor GWO 3600 

6.2  Electrode Maintenance 

If electrode can no longer be calibrated, it needs maintenance. 

Attention! The electrolyte is a corrosive. 

To maintain electrode please proceed as follows: 

1. Unscrew diaphragm head and wipe clean of electrolyte solution using a paper cloth. Do not touch 

electrolyte. If your skin had contact with electrolyte, immediately rinse thoroughly with clear water. 

2. Clean silver cathode with sand paper (grain size 240). Do not polish silver cathode, surface should stay 

rough. Remove all dust. 

3. Remove filling screw and top up lost electrolyte (e.g. using disposable syringe)) 
4. Put back and tighten filling screw. 
5. Top up diaphragm head with electrolyte avoiding air bubbles and place on table (cover table with 

absorbent paper first). 

6. Keep electrode in a vertical position and screw diaphragm head to the electrode from the bottom. 

Electrolyte will be forced out of the diaphragm head and spill over (put on disposable gloves or use 
paper towel to touch diaphragm head). 

7. Wipe up excess electrolyte with paper cloth. 

 

 

 

 

 

 

The oxygen electrode is an active electrode consisting of a silver cathode 
and a lead anode with the electrolyte being potassium hydroxide (KOH). In 
case of oxygen being present it will be reduced at the silver cathode, i.e. 
the electrode supplies a current. No oxygen means no current either. The 
oxygen measurements use up both the silver cathode as well as the lead 
anode. The electrode is subject to ageing. Therefore, we recommend to 
maintain the electrode at monthly intervals (p.r.t. 'Electrode maintenance') 

 

 

Please note: Make it a rule to always store the electrode in a 
humid environment. 
- in the storage flask filled with water 
- in another container filled with water 
If electrode has not been used for some time, clean diaphragm 
with soft cloth and remove deposits, if any (algae, bacteria 
etc.). 

 

Electrode design

 

 

The electrode housing is made of PVC. With the exception of the electrode 
shaft all parts need to be maintained regularly and be replaced if 
necessary. 

o  Protective flask

: The protective flask is used to moisten the 

diaphragm. This The prolongs service life of the electrode. The 
protective flask contains water. Attention! Use water only; never use 
potassium chloride (KCI); this is only required for storage of pH-
electrode. 

o Diaphragm head

: the diaphragm head is covered with a teflon 

diaphragm. It will be filled with KOH electrolyte and screwed onto the 
electrode shaft (no air bubbles). Damages in the diaphragm, large air 
bubbles or air bubble rings in the diaphragm head will result in 
erroneous measurements. This may also be the reason for errors in 
the calibration. The diaphragm head is a spare part and can be 
ordered individually. 

o Filling hole

: If the electrode is used at high temperatures or if it has 

been stored without its protective flask for a longer period of time, 
some electrolyte will be lost due to evaporation. During maintenance 
make it a rule to unscrew diaphragm head, remove locking screws 
and top up electrolyte using a syringe. Replace and tighten locking 
screws. 

 

 

 
 

Attention when working with electrolyte! The electrolyte is 
corrosive! (strong base, KOH) 

 

 

 

 

 

Filling hole

Electrode shank

O-ring

Diaphragm cap

Summary of Contents for GMH 3651

Page 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 0 Operating Manual GMH 3651 Oxygen meter for dissolved oxygen with integrated temperature and pressure...

Page 2: ...IBRATION AL 1 PT 11 8 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 12 8 3 EVALUATION OF SENSOR STATE ELE 12 9 OPERATION OF THE LOGGER 13 9 1 FUNC STOR STORING SINGLE MEASUREMENT SETS 13 9 2 FUNC CYCL AUTOM...

Page 3: ...s device has been designed and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safet...

Page 4: ...s to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V power supplies often hav...

Page 5: ...4 ok arrow Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 SAL arrow Signals that the salinity correction is aktivated 7 Logg arrow Sh...

Page 6: ...lip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attached to a belt etc The device with pop up clip at position 90 can b...

Page 7: ...h in meter only with depth accessory TMV 3600 DEP Depth measuring activated Second display always shows depth in meter only with depth accessory TMV 3600 U N T T C All temperatures in degree Celsius e...

Page 8: ...rement 0 0 SET AL Set Alarm AL 1 ON NO SO Monitoring Oxygen Alarm on with buzzer Alarm on without buzzer OFF No alarm monitoring for oxygen AL N ONC Monitoring Oxygen Concentration in P 02 Monitoring...

Page 9: ...to ageing Therefore we recommend to maintain the electrode at monthly intervals p r t Electrode maintenance Please note Make it a rule to always store the electrode in a humid environment in the stora...

Page 10: ...xygen concentration requires entering the salinity of the medium p r t Configuration The correction of salinity is based on media on a water basis whose chemical content is similar to sea water The co...

Page 11: ...ctrode is exposed to air with a relative atmospheric humidity of 100 Proceed as follows Put some distilled water in a bottle Close bottle and generate a water steam saturation 100 rel atmospheric humi...

Page 12: ...blinks no valid reference is recognised by the instrument As soon the values for oxygen and temperature are stable the calibration of the second point will be finished At 2 point calibration the calib...

Page 13: ...f data sets 1000 A data set contains oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute p...

Page 14: ...The recording automatically will be stopped Stopping the recording manually By pressing Store key 6 the recording can be stopped manually Then the following choice appears The selection can be made by...

Page 15: ...nels oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute pressure in hPa abs or mmHg abs o...

Page 16: ...rtly switch between ELE total electrode state evaluation in SL 1 Slope at 0 Air saturated water 1 SL 2 Slope at air saturated water 100 1 Date time display of data set or Change between the different...

Page 17: ...or Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery will be exhausted soon Further measurements a...

Page 18: ...000 data sets with measuring point input Alarm Buzzer visual interface 2 channels selectable oxygen unit and temperature Additional functions Min max hold auto hold Housing Break proof ABS housing Pro...

Reviews: