background image

H74.0.05.6C-06 

Operating Manual GMH 3651 

page 16 of 18 

12 GLP 

GLP (Good Laboratory Practice) includes regular check of devices and accessories. For oxygen 
measurements it is highly important to ensure correct pH calibration. The device provides the following 
functions to help with this. 
The usage of the GLP functions is only reasonable if the sensor is not changed. Although all data is stored in 
the device, it refers to the particular sensor. 

12.1  Calibration interval (

[.INT

You can enter the interval after which the device reminds you to recalibrate in the configuration.  
The interval times should be chosen according to the application and the stability of the sensor.  
“CAL” flashes on the display as soon as the interval has expired.  

12.2  Calibration storage (rEAd CAL)  

The last calibrations are stored with results and date and can be read out. 

Display calibration data: 

Historical calibration data can be comfortably read out via PC software GMHKonfig and GSOFT3050 or 
displayed directly at the device:

 

 

 

 

Press for 2 seconds

  

The display will show:

 

 

or 

 

(configuration level)

 

 

Press several times until this is 
displayed:

 

 

read cal. = “read calibration data” 

 

Press shortly:

 switch between  

ELE[

 = total electrode state evaluation in % 

SL. 1

 = Slope at 0% ... Air saturated water *

1)

 

SL. 2

 = Slope at air saturated water ... 100% *

1)

 

Date + time display of data set  

 or 

 

Change between the different calibration data sets 

 

 

Abort calibration storage display 

*

1

at 1 and 2-point calibration is SL.1 = SL. 2 

   at 3-point calibration there are different slopes for the referring range segments. 

13 Alarm 

(„AL.“) 

There are three possible settings: 
Alarm off (AL. oFF), on with buzzer (AL. on), on without buzzer (AL. no.So). 
Following conditions will display an alarm, when the function is activated (on or no.So): 

-

 

Value is below lower (AL. Lo) or above upper alarm rail (AL.Hi). 

-

 

Sensor error  

-

 

Low battery (bAt) 

-

 

Err.7: System error (always with buzzer) 

In case of an alarm and when polling the interface the “prio”-flag is set in the returned message. 

14 

Real Time Clock („CLOC“) 

The real time clock is used for the logger and GLP  functions: Recorded values are also containing the point 
of time, when they were measured. Please check the settings when necessary. 
If the battery was replaced the referring menu ‚CLOC‘ will automatically be started 

 

 

1

2

3

4

5

6

ON
OFF

max

Store

CAL

min

Set

Menu

Quit

1

2

3

4

5

6

ON
OFF

max

Store

CAL

min

Set

Menu

Quit

1

2

3

4

5

6

ON
OFF

max

Store

CAL

min

Set

Menu

Quit

1

2

3

4

5

6

ON
OFF

max

Store

CAL

min

Set

Menu

Quit

1

2

3

4

5

6

ON
OFF

max

Store

CAL

min

Set

Menu

Quit

Summary of Contents for GMH 3651

Page 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 0 Operating Manual GMH 3651 Oxygen meter for dissolved oxygen with integrated temperature and pressure...

Page 2: ...IBRATION AL 1 PT 11 8 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 12 8 3 EVALUATION OF SENSOR STATE ELE 12 9 OPERATION OF THE LOGGER 13 9 1 FUNC STOR STORING SINGLE MEASUREMENT SETS 13 9 2 FUNC CYCL AUTOM...

Page 3: ...s device has been designed and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safet...

Page 4: ...s to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V power supplies often hav...

Page 5: ...4 ok arrow Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 SAL arrow Signals that the salinity correction is aktivated 7 Logg arrow Sh...

Page 6: ...lip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attached to a belt etc The device with pop up clip at position 90 can b...

Page 7: ...h in meter only with depth accessory TMV 3600 DEP Depth measuring activated Second display always shows depth in meter only with depth accessory TMV 3600 U N T T C All temperatures in degree Celsius e...

Page 8: ...rement 0 0 SET AL Set Alarm AL 1 ON NO SO Monitoring Oxygen Alarm on with buzzer Alarm on without buzzer OFF No alarm monitoring for oxygen AL N ONC Monitoring Oxygen Concentration in P 02 Monitoring...

Page 9: ...to ageing Therefore we recommend to maintain the electrode at monthly intervals p r t Electrode maintenance Please note Make it a rule to always store the electrode in a humid environment in the stora...

Page 10: ...xygen concentration requires entering the salinity of the medium p r t Configuration The correction of salinity is based on media on a water basis whose chemical content is similar to sea water The co...

Page 11: ...ctrode is exposed to air with a relative atmospheric humidity of 100 Proceed as follows Put some distilled water in a bottle Close bottle and generate a water steam saturation 100 rel atmospheric humi...

Page 12: ...blinks no valid reference is recognised by the instrument As soon the values for oxygen and temperature are stable the calibration of the second point will be finished At 2 point calibration the calib...

Page 13: ...f data sets 1000 A data set contains oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute p...

Page 14: ...The recording automatically will be stopped Stopping the recording manually By pressing Store key 6 the recording can be stopped manually Then the following choice appears The selection can be made by...

Page 15: ...nels oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute pressure in hPa abs or mmHg abs o...

Page 16: ...rtly switch between ELE total electrode state evaluation in SL 1 Slope at 0 Air saturated water 1 SL 2 Slope at air saturated water 100 1 Date time display of data set or Change between the different...

Page 17: ...or Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery will be exhausted soon Further measurements a...

Page 18: ...000 data sets with measuring point input Alarm Buzzer visual interface 2 channels selectable oxygen unit and temperature Additional functions Min max hold auto hold Housing Break proof ABS housing Pro...

Reviews: