background image

H74.0.05.6C-06 

Operating Manual GMH 3651 

page 10 of 18 

8. Check cathode for air bubbles. 

If there are large air bubbles, remove diaphragm head again and repeat process as of point 5. If O-ring 
has been damaged, it has to be replaced. 

When maintenance has been completed plug on protective flask. Re-connect electrode to measuring device 
and wait for at least 3 hours till electrode can be calibrated.. 

Oxygen Measuring - please note 

Please observe the following points when measuring dissolved oxygen: 

  

- For measuring remove the protective flask. 

  

- Do not disconnect electrode from device

.  

  

 

If electrode has been disconnected, wait 2 - 3 hours till the final electrode signal has settled 

 

 

before carrying out measurements or a calibration. 

  

- Electrode needs to be calibrated 

(p.r.t. 'How to calibrate oxygen electrode') 

  

The temperatures of the electrode and of the liquid to be measured have to be identical  

  

 

(if necessary, wait till temperatures match) 

  

The Electrode has to  be submerged at least 3 cm into the liquid to being measured

 

  

The measured liquid has to stream along the electrode membrane with at least 30 cm/sec  

  

 

for measurements to be sufficiently accurate: either stir continuously or use agitator. 

- The electrode measurement is sensitive against shocks!  
  

By stirring of the electrode in the measured liquid be careful that the electrode does not hit the  

  

container. A vibration of the electrode has a effect to the measured value. 

-The 

optimum operation position 

is: with the sensor inlet pointing downwards 

The instrument calculates the oxygen concentration [mg/l], the oxygen saturation [%] and the oxygen partial 
pressure [hPa] from the electrode signal and the temperature. According to DIN38408-C22 all 
measurements refer to steam saturated air. 

7.1  Correction of salinity ('SAL') 

The higher the salinity (salt content) the lower the solubility of oxygen in water, i.e. although the partial 
oxygen pressure is the same, the quantity of oxygen dissolved in water (mg/l) is lower. Therefore, 
determination of the oxygen concentration requires entering the salinity of the medium (p.r.t. 'Configuration'). 
The correction of salinity is based on media on a water basis, whose chemical content is similar to sea water. 
The corrections are based on the 'International Oceanographic Tables' (IOT). 

7.2  Ambient pressure and measuring depth of the electrode 

The pressure at the sensor membrane is important for: 

  The calculation oft he oxygen saturation (%sat). 

At air water can get 100% saturation. Assumed that there are no oxygen consuming processes 
(biological degradation, chemical effects) and that there are no oxygen enriching processes (e.g. 
excessive ventilation or photosynthesis) 

  The calculation of oxygen concentration (mg/l or ppm) 
  The electrode evaluation at calibration 

Therefore it is necessary to compensate the pressure influence via integrated sensor or, like practised with 
more primitive instruments via tables and manual input of pressure or Elevation above sea level values. 

7.2.1  Measurement without Water depth measuring assembly: Menü dEPt = OFF 

The integrated pressure sensor is permanently measuring the ambient pressure to: 

  Compensate the pressure for the concentration [mg/l or ppm] or saturation [%O2] 
  Pressure compensate the calibration 

7.2.2  Measurement without Water depth measuring assembly  TMV 3600 (dEPt = On) 

Necessary accessory: Water depth measuring assembly TMV 3600. 
The pressure sensor measures the hydrostatic pressure to evaluate the measuring depth of the sensor with 
a resolution of 0.1m 
(e.g. for comfortable logging of oxygen profiles in lakes) 
Before starting the measuring (during segment test) and during calibration the pressure port has to be at 

ambient pressure: The reference ambient pressure is taken (display 

P.REF

 and pressure) and memorized for 

calculation of saturation [%O2], and as reference for the hydrostatic depth measuring. 
Info: The hydrostatic pressure does not really play a significant role for the calculation of concentration (mg/l 
or ppm) or saturation [%O2] up to 100m depth, but it is an important unit to evaluate oxygen layering in 
waters. Combined with the logger function evaluation can be performed very comfortably! 
More detailed Info: Please refer to manual of TMV 3600.

 

 

 

Summary of Contents for GMH 3651

Page 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 0 Operating Manual GMH 3651 Oxygen meter for dissolved oxygen with integrated temperature and pressure...

Page 2: ...IBRATION AL 1 PT 11 8 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 12 8 3 EVALUATION OF SENSOR STATE ELE 12 9 OPERATION OF THE LOGGER 13 9 1 FUNC STOR STORING SINGLE MEASUREMENT SETS 13 9 2 FUNC CYCL AUTOM...

Page 3: ...s device has been designed and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safet...

Page 4: ...s to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V power supplies often hav...

Page 5: ...4 ok arrow Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 SAL arrow Signals that the salinity correction is aktivated 7 Logg arrow Sh...

Page 6: ...lip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attached to a belt etc The device with pop up clip at position 90 can b...

Page 7: ...h in meter only with depth accessory TMV 3600 DEP Depth measuring activated Second display always shows depth in meter only with depth accessory TMV 3600 U N T T C All temperatures in degree Celsius e...

Page 8: ...rement 0 0 SET AL Set Alarm AL 1 ON NO SO Monitoring Oxygen Alarm on with buzzer Alarm on without buzzer OFF No alarm monitoring for oxygen AL N ONC Monitoring Oxygen Concentration in P 02 Monitoring...

Page 9: ...to ageing Therefore we recommend to maintain the electrode at monthly intervals p r t Electrode maintenance Please note Make it a rule to always store the electrode in a humid environment in the stora...

Page 10: ...xygen concentration requires entering the salinity of the medium p r t Configuration The correction of salinity is based on media on a water basis whose chemical content is similar to sea water The co...

Page 11: ...ctrode is exposed to air with a relative atmospheric humidity of 100 Proceed as follows Put some distilled water in a bottle Close bottle and generate a water steam saturation 100 rel atmospheric humi...

Page 12: ...blinks no valid reference is recognised by the instrument As soon the values for oxygen and temperature are stable the calibration of the second point will be finished At 2 point calibration the calib...

Page 13: ...f data sets 1000 A data set contains oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute p...

Page 14: ...The recording automatically will be stopped Stopping the recording manually By pressing Store key 6 the recording can be stopped manually Then the following choice appears The selection can be made by...

Page 15: ...nels oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute pressure in hPa abs or mmHg abs o...

Page 16: ...rtly switch between ELE total electrode state evaluation in SL 1 Slope at 0 Air saturated water 1 SL 2 Slope at air saturated water 100 1 Date time display of data set or Change between the different...

Page 17: ...or Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery will be exhausted soon Further measurements a...

Page 18: ...000 data sets with measuring point input Alarm Buzzer visual interface 2 channels selectable oxygen unit and temperature Additional functions Min max hold auto hold Housing Break proof ABS housing Pro...

Reviews: