9  Operation of the Logger 

The device supports two different logger functions: 
 

„Func-Stor“:

  

each time when „store“ (key 6) is pressed a measurement will be recorded. 

„Func-CYCL“:

  measurements will automatically be recorded at each interval, which was set in the logger 

menu ‚CYCL‘ until the logger will be stopped or the logger memory is full.  
The recording is started by pressing „Store“ 2 seconds. 

For the evaluation of the data the software GSOFT3050 (V3.5 or higher) has to be used. The software also 
allows easy configuration and starting of the logger. 

When the logger is activated (Func Stor or Func CYCL) the hold and auto hold functions are no longer 
available, key 6 is solely used for the operation of the logger functions. 

9.1   

„Func-Stor“: Storing Single Measurement Sets 

Each time when „store“ (key 6) is pressed a measurement and its time stamp will be recorded.  
The recorded data can be viewed either in the display (when calling the configuration an additional menu 
„REAd LoGG“ is displayed, see below) or by means of the interface and a PC with GSOFT3050-software. 

The logger stores the current measuring, independent from the stability of the value. 

Max. number of data sets: 

1000  

A data set contains: 

- oxygen concentration in mg/l or ppm 
- oxygen saturation in percent or partial pressure in hPa or mmHg 
- temperature value at the time of recording in °C or °F 
- absolute pressure in hPa or mmHg or depth in m  
- time and date of the recording 

After each recording „St. XX“ will be displayed for a short time. XX represents the number of the recording. 
 

Input of the measuring point 

“L-Id”: Selection of measuring point via keys 

 or 

Number 0…19999 or text assigned to number 1…40  
(Comfortable assignment of texts can be done with gratis software GMHKonfig)  
Confirm input with 

„store“. 

 

When logger memory contains recordings already: 

When „Store“ is pressed for 2 seconds, the choice for clearing the logger memory will be displayed: 

 
The selection can be made by 

 (key 2) and 

 (key 5). "Quit" (key 6) enters the choice. 

 

If the logger memory is full, the display will show: 

 

 

 
Viewing Recorded Data 

Within the „LoGG Stor“ function the logger data can be viewed directly in the display not only by means of a 
computer (like at „Func CYCL“): press 2 seconds „Set“ (key 4): The first menu displayed now is „rEAd LoGG“ 
(read logger data). After pressing 

 (key 3)  the measurement recorded last will be displayed, changing 

between the different data referring to the measurement also is done by pressing 

.  

Changing the measurement is done by pressing the keys 

 or 

 
 

 

Clear all 
recordings 

Clear the last 
recording  

Clear nothing  

(cancel menu)  

Summary of Contents for GMH 3651

Page 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 0 Operating Manual GMH 3651 Oxygen meter for dissolved oxygen with integrated temperature and pressure...

Page 2: ...IBRATION AL 1 PT 11 8 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 12 8 3 EVALUATION OF SENSOR STATE ELE 12 9 OPERATION OF THE LOGGER 13 9 1 FUNC STOR STORING SINGLE MEASUREMENT SETS 13 9 2 FUNC CYCL AUTOM...

Page 3: ...s device has been designed and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safet...

Page 4: ...s to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V power supplies often hav...

Page 5: ...4 ok arrow Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 SAL arrow Signals that the salinity correction is aktivated 7 Logg arrow Sh...

Page 6: ...lip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attached to a belt etc The device with pop up clip at position 90 can b...

Page 7: ...h in meter only with depth accessory TMV 3600 DEP Depth measuring activated Second display always shows depth in meter only with depth accessory TMV 3600 U N T T C All temperatures in degree Celsius e...

Page 8: ...rement 0 0 SET AL Set Alarm AL 1 ON NO SO Monitoring Oxygen Alarm on with buzzer Alarm on without buzzer OFF No alarm monitoring for oxygen AL N ONC Monitoring Oxygen Concentration in P 02 Monitoring...

Page 9: ...to ageing Therefore we recommend to maintain the electrode at monthly intervals p r t Electrode maintenance Please note Make it a rule to always store the electrode in a humid environment in the stora...

Page 10: ...xygen concentration requires entering the salinity of the medium p r t Configuration The correction of salinity is based on media on a water basis whose chemical content is similar to sea water The co...

Page 11: ...ctrode is exposed to air with a relative atmospheric humidity of 100 Proceed as follows Put some distilled water in a bottle Close bottle and generate a water steam saturation 100 rel atmospheric humi...

Page 12: ...blinks no valid reference is recognised by the instrument As soon the values for oxygen and temperature are stable the calibration of the second point will be finished At 2 point calibration the calib...

Page 13: ...f data sets 1000 A data set contains oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute p...

Page 14: ...The recording automatically will be stopped Stopping the recording manually By pressing Store key 6 the recording can be stopped manually Then the following choice appears The selection can be made by...

Page 15: ...nels oxygen concentration in mg l or ppm oxygen saturation in percent or partial pressure in hPa or mmHg temperature value at the time of recording in C or F absolute pressure in hPa abs or mmHg abs o...

Page 16: ...rtly switch between ELE total electrode state evaluation in SL 1 Slope at 0 Air saturated water 1 SL 2 Slope at air saturated water 100 1 Date time display of data set or Change between the different...

Page 17: ...or Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery will be exhausted soon Further measurements a...

Page 18: ...000 data sets with measuring point input Alarm Buzzer visual interface 2 channels selectable oxygen unit and temperature Additional functions Min max hold auto hold Housing Break proof ABS housing Pro...

Reviews: