For each of the TEOAE, DPOAE and ABR modalities the Novus was controlled from the
PC so that many repeat tests could be automatically performed. The PC was used to
control the level of noise, from the simulator, injected into the instrument so that it
experienced the full range of noise present in a typical clinical environment.
R
ESULTS
ABR mode: 525 tests were performed, no false ABR results occurred, the resulting
sensitivity has been assessed to be better than or equal to 99.6%
TEOAE mode: 137 tests were performed, no false OAE results occurred, the resulting
sensitivity has been assessed to be better than or equal to 99.6%
DPOAE mode: 370 tests were performed, no false OAE results occurred, the resulting
sensitivity has been assessed to be better than or equal to 99.6%