GSI Novus™ Newborn Hearing Screening System User Manual
D-0113050 Rev H
Page 75 of 96
PRINTER
Type/Model
Thermal printer MLP2, HM-E200
Connection
Wireless Bluetooth
®
Battery
Lithium Ion, DC 7.4V, 1500 mAh
Charger
100-250V AC, 50/60 Hz, 1.0 A
Weight
360 grams (12.7 oz.)
Paper
Thermal paper: 57.5 mm ± 0.5 mm (width)r
Labels: 57.5 mm ± 0.5 mm x 60 mm (width x length)
TEST SPECIFICATIONS
ABR
Test signals
CE-Chirp
®
(default) or Click,
Frequency range (200 Hz – 11 kHz), alternating polarity
Stimulus rate
88/sec left ear, 92.5/sec right ear
Stimulus level
35 dB nHL (default protocol), Range: 30-45 dB nHL
EEG filter
0.5 Hz – 5.0 kHz
Preamplifier Gain
72 dB
Input impedance
10 MΩ / 170 pF
Noise
<25 nV/
√
Hz
CMR Ratio
>100 dB at 100 Hz
Sample rate
22.05 kHz
A/D resolution
24 bit
Electrical impedance
measurement
Before and during testing
< 40 kΩ required for testing
Maximum test time
180 seconds (artifact free data samples)
DPOAE
Stimulus frequencies
2000, 3000, 4000, 5000 Hz (default protocol)
1500-6000 Hz range
Nominal frequency, F2/F1
ratio
F2, 1.22
Level L1/L2 (tolerance)
65/55 dB SPL ( ± 1.5 dB) – in ear calibration
A/D resolution
24 bit, 5.38 Hz resolution
SNR criteria minimum
6 dB (default protocol)
Response amplitude
minimum
-5 dB (default protocol)
Bands for Pass criteria
3 out of 4 (default protocol)
Analysis
Minimum 2 seconds to maximum 60 seconds
TEOAE
Center frequencies
1400, 2000, 2800, 4000 Hz (default protocol)
1000-4000 Hz range
Stimulus
Non-Linear click (IEC 60645-3) rate 71/sec