CHAPTER 4 Specification
External Input/Output Functions
Item
Description
Input/output types
Trigger input (1 ch) or External sampling input (1 ch)
Logic input (4 ch) or Pulse input (4 ch)
Alarm output (4ch), or Trigger output (1ch) + Alarm output (3ch)
* Trigger input and external sampling input can be switched.
* Alarm output and trigger output can be switched.
Input specifications
Maximum input voltage : 0 to +30V (single-ended ground input)
Input threshold voltage : Approx. +2.5V (logic input, pulse input)
Approx. +1.9V (external trigger, external sampling)
Hysteresis : Approx. 0.5V (+2.5V to +3.0V) (Logic input, pulse input)
Approx. 0.2V (+1.9V to + 2.1 V) (external trigger, external sampling)
Alarm output specifications
Output format: Open collector output (5 V, pull-up resistance 10 kΩ)
Contact capacity 5V to 24V, 100mA or less (0.2W or less)
Output conditions: Level judgment, window judgment, logic pattern judgment, pulse
judgment
Trigger output
When a trigger is detected after trigger output is set, a pulse of approx. 500 μs width is
output from the Output 1 terminal.
(Low active)
Pulse input
Pulse sampling interval: 10 μs to 1h
* Set it separately from sampling interval.
Setting faster than sampling interval is not possible (constant multiple)
Revolutions mode (engines, etc.)
Function: Mode to convert to the number of revolutions per minute by applying magnification
after counting the number of pulses for each pulse sampling interva
l.
Span: 10 to 500M
Counts mode (electric meters, etc.)
Function: Mode to accumulate the number of pulses for each pulse sampling interval from
the start of measurement
Span: Automatic (Maximum 20M)
Inst. mode
Function:
Mode to display the number of pulses for each pulse sampling interval.
The accumulated value for each pulse sampling interval is reset.
Span: 10 to 20M
Maximum number of pulse inputs
Maximum input frequency : 100 kHz
Maximum number of count : 15 MC (24-bit counter)
External sampling input
Maximum input frequency: 100 kHz
Temporal error: 1 μs or less.