RS–232 Operation
Micro-Ion E Module Instruction Manual 354011, Rev. A
31
Be
for
e
Yo
u Beg
in
Installa
tion
O
pe
ra
tio
n
Ove
rview
An
alog
O
p
e
ra
tio
n
M
ain
tena
nce
RS-
232
O
p
e
ra
tion
Read Micro-Ion E Gauge
ON/OFF Status
If you send the ion gauge status (IGS) command, the module returns a
character string that indicates the ON/OFF status of the Micro-Ion E gauge.
IGS command from host:
IGS
↵
Example response from module:
1 IG ON
↵
• The “1 IG ON” response means the Micro-Ion E gauge is ON.
• The “0 IG OFF” response means the Micro-Ion E gauge is OFF and the
module will return a “9.99E+09” error response to the RD (read pressure)
command.
The status LED indicates the operating status of the Micro-Ion E gauge.
• Figure 5-2 illustrates the status LED on the module.
• Table 5-7 lists conditions indicated by the status LED.
Figure 5-2
LED Status Indicator
Status LED
Table 5-7
Conditions Indicated by Status LED
LED Behavior
Condition Indicated
Solid green
Micro-Ion E gauge is ON. See page 30 to turn gauge OFF.
Solid amber
Micro-Ion E gauge is OFF. See page 30 to turn gauge ON.
Solid red
• Micro-Ion E gauge fault condition. See page 40 to replace gauge assembly.
• Overpressure shutdown. Decrease pressure to alleviate overpressure condition.
Summary of Contents for Micro-Ion 354 E Series
Page 2: ......
Page 10: ...10 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 24: ...24 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 42: ...Chapter 6 42 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 46: ...46 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 52: ...52 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 53: ......