Chapter 5
30
Micro-Ion E Module Instruction Manual 354011, Rev. A
Reset Baud Rate to
Factory Defaults
You may restore the baud rate to its default value of 9600 baud by sending
a factory reset (FAC) command.
FAC command from host:
FAC
↵
Response from module:
PROGM OK
↵
After sending the FAC command, you
must
send the reset (RST) command.
Reset command from host:
RST
↵
Response from module:
None
For more information about the RST command, see page 33.
Turn Micro-Ion E Gauge
ON or OFF
The Micro-Ion E gauge measures vacuum pressure for N
2
or air within the
ranges listed in Table 5-6. As pressure decreases to 5 x 10
–3
Torr
(6.66 x 10
–1
Pa, 6.66 x 10
–3
mbar), you
must
send the ion gauge on (IG1)
command to turn the Micro-Ion E gauge ON and enable pressure readings.
IG1 command from host:
IG1
↵
Response from module:
PROGM OK
↵
The command includes the alpha characters “I” and “G” and the numeric
value 1 (ON).
In some instances, you might need to turn the Micro-Ion E gauge OFF.
Send the ion gauge off (IG0) command to turn the Micro-Ion E gauge OFF
and disable pressure readings.
IG0 command from host:
IG0
↵
Response from module:
PROGM OK
↵
The command includes the alpha characters “I” and “G” and the numeric
value 0 (OFF).
Table 5-6
Micro-Ion E Gauge Pressure Measurement Range
Low Pressure Limit (N
2
or air)
High Pressure Limit (N
2
or air)
5 x 10
–7
Torr
5 x 10
–3
Torr
6.66 x 10
–5
Pa
6.66 x 10
–1
Pa
6.66 x 10
–7
mbar
6.66 x 10
–3
mbar
Summary of Contents for Micro-Ion 354 E Series
Page 2: ......
Page 10: ...10 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 24: ...24 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 42: ...Chapter 6 42 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 46: ...46 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 52: ...52 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 53: ......