02.11.2018
25
4.18 Set the Capture time without Capture
Transmit
Description
Receive
Example
Comment
setcaptimexyz b
x = set capture time
1=600 ms, 2=200ms, 3=120ms,
4=60ms, 5=20ms, 6=10ms,
7=2ms, 8= user programmable,
0=off (without capture - only set)
OK
setcaptime2115
OK
x=2 200ms
y=1 sensor low 9x9
z=1 sensor 1-5 or 1...495
b=board, if z=1-5
4.15 Set the X Chromaticity Offset
Transmit
Description
Receive
Example
Comment
setx-0.xxx
# = fiberno. 1..5 b
or # = 1..495
set x chromaticity offset
OK
setx0.050
OK
set fiber 1 x offset to
+0.050
#=sensor,
xxx=value,
b=board
4.16 Set the Y Chromaticity Offset
Transmit
Description
Receive
Example
Comment
sety-0.xxx
# = fiberno. 1..5 b
or # = 1..495
set Y chromaticity Off
OK
setyoffset1-0.050
OK
set fiber 1 Y offset to
-0.050
#=sensor,
xxx=value,
b=board
4.17 Get distance between LED and Board
Transmit
Description
Receive
Example
Comment
getdistance# b
# = fibernr.
1..5 oder # =
1..495, b =
board
Get stored distance value in mm xxx.x
getdistance4
003.5
# = 1..5 wenn b = board
# = 1..495 bei max. 99
boards
b not specified
Transmit
Description
Receive
Example
Comment
setaverage## b
average factor
0=5 … 15=80 average
(without capture - only set)
OK
setaverage10
##=0 … 15
b=board
4.19 Set the Average Factor without Capture