g. Remove the two (2) screws located next to the lower column connections inside the
column oven.
h. Remove the four (4) screws located on the top and side of the CONTROL UNIT and
remove the CONTROL UNIT COVER.
i. Follow the detector cable and disconnect the detector cable from the Molex connector
located in the CONTROL UNIT.
NOTE
)
Attach a string or wire to the Molex connector before pulling the
detector leads out. This will assist in reinstallation of the detector.
j. CAREFULLY remove the DETECTOR HOUSING. In doing so, CAREFULLY draw the
DETECTOR CABLE through the clearance hole from the CONTROL UNIT into the OVEN
UNIT.
k. Remove the DETECTOR HOUSING top by removing the four (4) screws.
l. Remove the DETECTOR HOUSING front panel by removing the four (4) screws.
m. Remove the DETECTOR HOUSING rear panel by removing the four (4) screws.
n. CAREFULLY lift the DETECTOR out of cell housing.
4. Filament Replacement
To replace fi laments, proceed as follows:
a. Place cell in a vise.
b. Using a 1/2” wrench, remove each of the four (4) bushings holding each of the four (4)
fi laments in place.
c. Examine the detector for any accumulation of material and general cleanliness.
d. Clean the detector if necessary before installing new fi laments.
e. Refer to the General Service Bulletin (following the WARRANTY page of this manual) for
recommendations of installing new fi laments.
f. The new fi laments should be installed in the detector, the bushings tightened, and a leak
check made BEFORE replacing the detector housing. Refer to the GENERAL SERVICE
BULLETIN for further assistance.
Summary of Contents for 580 series
Page 2: ...GOW MAC Instrument Co 2 ...
Page 6: ...GOW MAC Instrument Co 6 ...
Page 8: ...GOW MAC Instrument Co 8 ...
Page 10: ...GOW MAC Instrument Co 10 ...
Page 24: ...GOW MAC Instrument Co 24 ...
Page 36: ...GOW MAC Instrument Co 36 ...
Page 42: ...GOW MAC Instrument Co 42 ...
Page 54: ...GOW MAC Instrument Co 54 ...
Page 62: ...GOW MAC Instrument Co 62 ...
Page 66: ...GOW MAC Instrument Co 66 ...
Page 68: ...GOW MAC Instrument Co 68 ...