Series 580 TCD Gas Chromatograph
September 2017, Rev. 5
41
Inj.
Series
Bypass
45 secretary
.
4 min
Series
H
2
O
x1
H
2
x1
CO
x1
C
2
H
6
x1
O
2
x8
CO
2
x1
CH
4
x8
N
2
x8
Typical Chromatogram
Analysis of Furnace Gases
Instrument:
Series 580
Detector:
TCD
Column:
See Note A
Carrier Gas:
Helium
Carrier Flow:
40 cc/min.
Temperature:
Column
59 °C
Detector
100 °C
Bridge Current:
150 mA
Attenuation:
Noted
Sample:
Gas Mix
Sample Size:
2 cc
Note A:
Column #1 – 7’ x 1/8” Porapak Q, 80/100 mesh
Column #2 – 8’ x 1/8” Molecular Sieve 5A, 80/100 mesh