Symptom
Cause
Remedy
Constant baseline drift in
Detector temperature
Allow suffi cient time for
one direction.
increasing (decreasing).
detector to restabilize after
changing its temperature.
Leaks causing fi lament
Check for leaks.
oxidation.
Column bleed.
Replace column(s).
Irregular baseline shifting.
Column not properly
Condition column.
conditioned.
Excessive column bleeding
Use diff erent column. Some
from well conditioned column.
packing materials cannot be
operated at elevated
temperatures without diffi culty.
This symptom may occur even
on well conditioned columns in
which carrier gas fl ow rates
have been carefully optimized.
High background noise
Column contaminated
Recondition column.
or excessive column bleed.
Contaminated injector.
Clean injector and replace
septum.
Contaminated detector.
Clean detector.
Defective recorder.
Set GC Attenuator to infi nity
(
). If noise continues, check
recorder. See recorder
manual.
Recorder slidewire dirty.
Clean recorder slidewire.
Carrier gas fl ow rate
Reduce carrier gas fl ow rate.
too high.
Carrier gas leaking.
Locate and correct leak.
Bad ground connection.
Make sure all instruments are
connected to good earth
ground.
Electronic circuitry dirty.
Consult GOW-MAC
Engineering.
Summary of Contents for 580 series
Page 2: ...GOW MAC Instrument Co 2 ...
Page 6: ...GOW MAC Instrument Co 6 ...
Page 8: ...GOW MAC Instrument Co 8 ...
Page 10: ...GOW MAC Instrument Co 10 ...
Page 24: ...GOW MAC Instrument Co 24 ...
Page 36: ...GOW MAC Instrument Co 36 ...
Page 42: ...GOW MAC Instrument Co 42 ...
Page 54: ...GOW MAC Instrument Co 54 ...
Page 62: ...GOW MAC Instrument Co 62 ...
Page 66: ...GOW MAC Instrument Co 66 ...
Page 68: ...GOW MAC Instrument Co 68 ...