GOW-MAC Instrument Co.
18
3. Fill out and mail the yellow WARRANTY-REGISTRATION CARD (included with this manual)
to ensure that the warranty will be validated and that you will be kept informed of any
improvements or other items of interest.
D. Location
1. The Series 580 GC should be placed in a location that is secure, vibration-free, protected
from abrupt temperature changes (maximum ambient temperature range is 15 °C to 40 °C),
and drafts. Such changes may upset the temperature stability in the course of an analysis or
preparation.
2. Enough adjacent tabletop space should be allowed for the installation of recorders,
integrators, computers, etc. Allow suffi cient space on all sides of the Series 580 GC for easy
access.
3. Make sure there is adequate space for the installation of the carrier gas cylinder. Cylinders
should be securely fastened to the wall or table.
CAUTION
READ “SECTION 2 - SAFETY” TO ENSURE PROPER HANDLING
OF GAS CYLINDERS.
4. An electrical outlet (ac) should be near the location where the Series 580 GC is to be
installed. If the outlet is not a 3-pin type, make sure that a good ground connection is
available, since a good ground is necessary for proper operation. The ac outlet should be
connected to a circuit that is not heavily loaded with other electrical equipment because input
voltage to the instrument should be steady for optimum operating stability.
If the ac line voltage varies, consideration should be given to the installation of a stabilizing
transformer at the ac outlet.
NOTE
)
Both recorder and the Series 580 GC should be connected to the
same duplex service outlet to prevent ground loops.
E. Power Requirements
The Series 580 GC requires a 115 volt/60 Hz power source capable of providing up to 10 amps.
The Series 582 GC requires a 230 volt/50 Hz power source capable of providing up to 5 amps.
The ac power cord is terminated with a straight-blade 3-prong plug rated for 15 amp service that
requires a matching receptacle.
CAUTION
MAKE SURE ALL SWITCHES ON FRONT AND BACK OF THE
SERIES 580 GC ARE IN THE “OFF” POSITION BEFORE PLUGGING
IN THE INSTRUMENT.
Summary of Contents for 580 series
Page 2: ...GOW MAC Instrument Co 2 ...
Page 6: ...GOW MAC Instrument Co 6 ...
Page 8: ...GOW MAC Instrument Co 8 ...
Page 10: ...GOW MAC Instrument Co 10 ...
Page 24: ...GOW MAC Instrument Co 24 ...
Page 36: ...GOW MAC Instrument Co 36 ...
Page 42: ...GOW MAC Instrument Co 42 ...
Page 54: ...GOW MAC Instrument Co 54 ...
Page 62: ...GOW MAC Instrument Co 62 ...
Page 66: ...GOW MAC Instrument Co 66 ...
Page 68: ...GOW MAC Instrument Co 68 ...