6-1
TEST / CALIBRATION EXAMPLE
TEST / CALIBRATION SETUP OF
GMI PORTABLE INSTRUMENT
The following procedure details how to setup the test / calibration
file for a GMI portable instrument using GMI
flexiCal
net
software.
6.1 CALIBRATE GASURVEYOR 526
INSTRUMENT
In this example, calibration is to be performed on a GMI
Gasurveyor 526 instrument.
To perform calibration, the calibration file for the instrument must
be sent to a GDU
net
.
To setup the Gasurveyor 526
calibration file, ensure
flexiCal
net
has been installed and the GDU
net
has been registered on the
IMS server. Refer to chapters 2 and 3 for details.
6.1.1
Start Software
To start the
flexiCal
net
software package:
Select Start / All Programs / Gas Measurement Instruments Ltd
/ flexiCal net or, double click on
flexiCal net
the desktop.
The window is illustrated in Fig. 6-1 is now displayed: