5.3.1 Test / Calibration Procedure
The following procedure will allow correct operation for testing /
calibrating an instrument.
1. Ensure correct hardware installation has taken place, refer to
the GDU
net
Installation Guide, part no. 14766.
2. Ensure
flexiCal
net
software installation has taken place,
refer to chapter 2 -
flexiCal
net
Installation.
3.
Send the required test / calibration files to the GDU
net
, refer
to chapter 3 -
flexiCal
net
Operation. If using the GDU
net
in standalone mode refer to chapter 7 - Standalone Mode.
4. Ensure
IMS Results Manager
installation and setup has
taken place, refer to chapter 4 - IMS Results Manager.
5. The procedure for the testing or calibrating of an instrument
is illustrated in Fig. 5-12. The test procedure shows each
step taken during the testing of an instrument from inserting
onto the GDU
net
until the removal of the instrument. The test
procedure for the keypad version of the GDU
net
is illustrated
in Fig. 5-13.
Note:
The test procedures also form the quick
operation guide shipped with the GDU
net
, part
no. 14768.
Warning: The GDU
net
may reset if there is an interruption
in the communications between the GDU
net
and
instrument under test. This interruption may be
due to the instrument battery power being low,
dirt in the optical communication windows, or
electromagnetic interference.