ACCESSORIES SUPPLIES 

DESCRIPTION 

CODE 

COMPLETE ASPIRATION JAR 1000ml 

RE 210001/02 

CONICAL FITTING 

RE 210420 

TUBES SET 8 mm x 14 mm 

51100/01 

HYDROPHOBIC AND ANTIBACTERIAL FILTER 

SP 0121 

SUCTION PROBE CH20 

25723 

AC/DC ADAPTER (UE60-140429SPA1) 

SP 0208/01 

POWER SUPPLY CORD FOR AC/DC ADAPTER 

SP 0020/03 

CIGARETTE LIGTHER CABLE 

SP 0007/02 

The filter is produced with (PTFE) hydrophobic material to prevent fluids entering the pneumatic circuit. It should be changed 
immediately if it becomes wet or if there is any sign of contamination or discolouration. If should also be changed if the unit is used 

with a patient whose risk of contamination is unknown. 

Don’t use the suction unit without the protection filter

. If the suction 

unit is used in an emergency or in a patient where the risk of contamination is not know the filter must be changed after each use.  

Available under request with different versions with complete jar 2000ml.  

Suction catheter

: Single-use device to be used on a single patient. Do not  wash or re-sterilize after use. Reuse may cause cross-

infections. Don’t use after lapse of the sell-by date 

Check the expiry date on the original packaging of the suction catheter and check the integrity of the sterile 
packaging. 

GIMA

 declines any liability for injury to the patient correlated to the deterioration of the above-

mentioned sterile packaging due to handling of the original packaging by third parties. 

WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 

10993-1 standards on material biocompatibility. 

Aspiration jar: The mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and sterilization.  
Beyond this limit, the physical-chemical characteristics of the plastic material may show signs of decay. Therefore, we recommend 

that you to change it. 

Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the said tube.  

Therefore, after each cleaning cycle, it is up to the final user to verify whether the tube is suitable for reuse. The component must 
be replaced if there are visible signs of decay of the material constituting the said component. 
Conical fitting: the number of cycles of sterilization and the number of cleaning cycles is strictly linked to the employment of the 

said component. Therefore, after each cleaning cycle, it is up to the final user to verify whether the fitting is suitable for reuse. 
The component must be replaced if there are visible signs of decay in the material constituting the said component. 
Service life of the device: More than 30000 hours of operation in accordance with the standard conditions of testing and operation. 

Shelf life: maximum 5 years from the date of manufactured. 

CLEANING OF ACCESSORIES 

Before using the device, the manufacturer advises you to clean and/or sterilize the accessories.

 

 

Washing and / or cleaning the autoclavable jar as to be carried out as follows: 

Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating substances;

Disconnect the tank from the device and remove the said container from the support of the device.

Separate all the parts of the cover (overflow device, washer).

Disconnect all tubes from the jar and the protection filter

Wash each part of the container from secretions under cold running water and then clean every single part in hot water 

(temperature not exceeding 60°C) 

Once again, carefully wash each single part using, if necessary, a non-abrasive brush to remove any deposits. 

Rinse with hot running water and dry all parts with a soft cloth (non-abrasive). It is possible to wash with commercial
disinfectants by carefully following the instructions and dilution values supplied by the manufacturer. 

After cleaning, leave the parts to dry in an open, clean environment. 

Dispose of the aspiration catheter according to that provided by local laws and regulations. 

The silicone aspiration tubes and the conical fitting may be carefully washed in hot water (temperature must not exceed 

60°C). After cleaning, leave the parts to dry in an open, clean environment.  

When cleaning is complete, reassemble the container for liquid aspirations according to the following procedure: 

Place the overflow valve into its seat in the cover (under VACUUM connector)

Insert floating valve keeping the o-ring towards the opening of the cage

Place the o-ring into its seat around the cover

After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leackages or liquid 
exit 

Summary of Contents for SUPERVEGA EVO BATTERY ASPIRATOR

Page 1: ...0 Gessate MI Italy gima gimaitaly com export gimaitaly com www gimaitaly com PROFESSIONAL MEDICAL PRODUCTS Gima S p A Via Marconi 1 20060 Gessate MI Italy Made in Italy Instruction manual SUPERVEGA EVO BATTERY ASPIRATOR 230 V 28183 ...

Page 2: ...r immerge the appliance into water Do not place or store the aspirator in places where it may fall or be pulled into the bathtub or washbasin In the event it is accidentally dropped do not attempt to remove the device from the water whilst the plug is still connected disconnect the mains switch remove the plug from the power supply and contact the GIMA technical service department Do not attempt t...

Page 3: ...ormatives IMPORTANT INFORMATION FOR CORRECT DISPOSAL OF THE PRODUCT IN ACCORDANCE WITH EC DIRECTIVE 2012 19 UE RAEE The symbol on the device indicates the separated collection of electric and electronic equipment At the end of life of the device don t dispose it as mixed solid municipal waste but dispose it referring to a specific collection centre located in your area or returring it to the distr...

Page 4: ...t the battery provided with this product shall not be treated as household waste By ensuring these batteries are disposed of correctly you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery The recycling of the materials will help to conserve natural resources At the end of their li...

Page 5: ...ting the said component Conical fitting the number of cycles of sterilization and the number of cleaning cycles is strictly linked to the employment of the said component Therefore after each cleaning cycle it is up to the final user to verify whether the fitting is suitable for reuse The component must be replaced if there are visible signs of decay in the material constituting the said component...

Page 6: ...evice is equipped with a Lithium Ion Battery which cannot be reached from the outside Only refer to GIMA technical service personnel to replace it The device should be checked at least once every 12 months by technical service Every 24 months it is compulsory to have a safety inspection and technical maintenance performed on the device ONLY USE BATTERIES RECOMMENDED BY GIMA USE OF BATTERIES OTHER ...

Page 7: ...ng adapter to the device with the appropriate connector and insert the power cable plug to the power socket To start the treatment press the ON OFF button and check its green back light When the device is switched on the LCD screen activates displaying a bar graph graduated scale and negative pressure value which must be set by the operator The negative pressure is set with keys and which are foun...

Page 8: ... cigarette lighter DC 12V Connect the device s external plug 12V to the ligther plug with the cigarette lighter cable Check the battery power status of the vehicle before the cigarette lighter cable Press the switch to start suction Press the switch to the I position to turn it on WARNING Only use the originally supplied or recommended replacement cigarette ligther cables view the chapter Importan...

Page 9: ...ed Battery operation guaranteed Fixed yellow back light During battery operation Intermediate status Battery not fully charged Battery operation guaranteed start the charge cycle upon activation of the red back light Fixed red back light During battery operation Low battery Start charge phase ATTENTION A long and continuous beep is produced during this signal to warn the operator about the low cha...

Page 10: ...he surgical SUPERVEGA EVO BATTERY ASPIRATOR is intended for use in the electromagnetic environment specifiedbelow The customers or the user of the surgical SUPERVEGA EVO BATTERY ASPIRATOR should assure that it s used in such an environment Emissions test Compliance Electromagnetic environment guidance Irradiated Conducted emissions CISPR11 Group 1 The surgical SUPERVEGA EVO BATTERY ASPIRATOR only ...

Page 11: ...itters and TV transmitters can not be theoretically and accurately foreseen To establish an electro magnetic environment generated by fixed RF transmitters an electro magnetic study of the site should be considered If the field intensity measured in the place where the device will be used surpasses the above mentioned applicable level of conformity the normal functioning of the device should be mo...

Page 12: ...not carried out by technical service RULES FOR RETURNING AND RAPAIRING UNDER NEW EUROPEAN RULES GIMA REQUIRES THE FOLLOWING PROCEDURES TO BECARRIEDOUT TO PROTECT THE INSTRUMENT ANDTHE SAFETY OF ALL WHO COME IN CONTACT WITH IT Before returning an instrument for repair the external surfaces and all accessories MUST be carefully disinfected with a cloth soaked in methylated spirits or hypochlorite ba...

Page 13: ...Gima S p A Via Marconi 1 20060 Gessate MI Italy gima gimaitaly com export gimaitaly com www gimaitaly com PROFESSIONAL MEDICAL PRODUCTS Gima S p A Via Marconi 1 20060 Gessate MI Italy ...

Reviews: