After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thouroughly. 

Then soak in warm water (temperature shall not exceed 60°C). Wash thouroughly and if necessary use a non-abrasive brush to 
remove incrustations. Rinse in running warm water and dry all parts with a soft cloth (non-abrasive). 

The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C 
(1 bar relative pressure – 15 min) making sure that the jar is positioned upsidedown. Mechanical resistance of the jar is 

guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1). Beyond this limit the 
physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore recommended. 

After sterilization and cooling at environment temperature of the parts make sure that these are not damaged. 
The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min). 

The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min).   

DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER 

PERIODICAL MAINTENANCE CHECKS 

The 

SUPERVEGA EVO BATTERY ASPIRATOR 

suction equipment does not need maintenance or lubrication.  

It is, however to inspect the unit before each use. With regard to training, given the information contained in the user manual and 
since it is easy to understand the said device, it doesn’t appear to be necessary. Unpack the instrument and 

always check

 integrity 

of plastic parts, they might have been damaged during previous use.  
Turn switch on and verify that loud noises are not present, these can indicate wrong functioning.  
Make sure you do not hear excessively annoying noises that may highlight a malfunction.  
The device is protected by a safety fuse (

F 10A L 250V

) situated in the cigarette lighter cable. When replacing, always check the 

type and value as indicated. 
Internally, the device is protected (see electrical specifications) by two fuses (

T 15A L 125V

) that cannot be reached from the 

outside. Therefore, contact the manufacturer to request the assistance of an authorized and qualified technician when they need to 
be replaced. If it’s replaced make sure that its replacement is always the same type and value, as indicated.  
The device is equipped with a Lithium-Ion Battery, which cannot be reached from the outside. Only refer to 

GIMA

 technical service 

personnel to replace it.  The device should be checked at least once every 12 months by technical service. Every 24 months it is 
compulsory to have a safety inspection and technical maintenance performed on the device. 

ONLY USE BATTERIES RECOMMENDED BY 

GIMA.

 USE OF BATTERIES OTHER THAN THOSE RECOMMENDED 

WILL MAKE THE WARRANTY VOID AND NULL. 
ONLY QUALIFIED PERSONNEL ARE AUTHORISED TO REPLACE THE INTERNAL BATTERY.  ANY OPERATION 
PERFORMED BY NON-TRAINED PERSONNEL MAY CAUSE DANGER (E.G. EXCESSIVE TEMPERATURE) 

Fault type

 

Cause

 

Solution

 

1.

 

Fixed Red Back-Light 

Low battery 

If it is intended to be used in ambulances, position the 

device on the support bracket and leave the battery charge 

until the Fixed White Back-Light gives the signal (TAB. I). 

2.

 

No LED is On and the Back-
Light failed to switch on 

Locking device 

Internal technical problem. Contact technical assistance. 

3.

 

No aspiration

Jar Cap badly screwed down 

Unscrewed the cap, then rescrew it correctly 

4.

 

No aspiration

Lid seal not in its seat 

Unscrew the cap and insert the seal properly in its seat 

5.

 

The float doesn’t close

If the cap has been washed, 

ensure that the float is not 

partially detached 

Insert the float into it’s place 

6.

 

The float doesn’t close

The float it’s covered by dirty 

material 

Unscrewed the cap, leave the and put in on autoclave 

7.

 

Low suction 

Foam inside the jar 

Fill the jar to 1/3 full of ordinary water  

8.

 

No aspiration due to flow

leakage of mucus 

Filter blocked 

Replace filter 

9.

 

The Vacuum power on the 
patient side is either very low

or absent 

 

Protection filter blocked or

damaged 

 

Connection tubes blocked, 

kinked or disconnected 

 

Shut-off valve blocked or

damaged 

 

Pump motor damaged

 

Replace the filter

 

Replace or reconnect the tubes, check the jar
connections 

 

Empty the jar, or disconnect the tube from the jar and 
unblock the shut-off valve. The unit twill only work in

the upright position 

 

Contact the technical service

10.

 

Noisy

Technical internal problem 

Contcat the technical service 

Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 – 

9 - 10 

None of the remedies has 

achieved the desired results 

Contact the seller or 

GIMA 

After-sales Assistance 

Service 

Lithium-Ion Battery Charging Cycles:

 The Lithium-Ion battery contained inside the device is guaranteed for a number greater 

than 300 charging cycles. In the vicinity of 300 charging cycles you can ask for verification of operating status to the manufacturer 

or to require replacement of the battery pack in such a way that you always have the component in perfect condition. 

Summary of Contents for SUPERVEGA EVO BATTERY ASPIRATOR

Page 1: ...0 Gessate MI Italy gima gimaitaly com export gimaitaly com www gimaitaly com PROFESSIONAL MEDICAL PRODUCTS Gima S p A Via Marconi 1 20060 Gessate MI Italy Made in Italy Instruction manual SUPERVEGA EVO BATTERY ASPIRATOR 230 V 28183 ...

Page 2: ...r immerge the appliance into water Do not place or store the aspirator in places where it may fall or be pulled into the bathtub or washbasin In the event it is accidentally dropped do not attempt to remove the device from the water whilst the plug is still connected disconnect the mains switch remove the plug from the power supply and contact the GIMA technical service department Do not attempt t...

Page 3: ...ormatives IMPORTANT INFORMATION FOR CORRECT DISPOSAL OF THE PRODUCT IN ACCORDANCE WITH EC DIRECTIVE 2012 19 UE RAEE The symbol on the device indicates the separated collection of electric and electronic equipment At the end of life of the device don t dispose it as mixed solid municipal waste but dispose it referring to a specific collection centre located in your area or returring it to the distr...

Page 4: ...t the battery provided with this product shall not be treated as household waste By ensuring these batteries are disposed of correctly you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery The recycling of the materials will help to conserve natural resources At the end of their li...

Page 5: ...ting the said component Conical fitting the number of cycles of sterilization and the number of cleaning cycles is strictly linked to the employment of the said component Therefore after each cleaning cycle it is up to the final user to verify whether the fitting is suitable for reuse The component must be replaced if there are visible signs of decay in the material constituting the said component...

Page 6: ...evice is equipped with a Lithium Ion Battery which cannot be reached from the outside Only refer to GIMA technical service personnel to replace it The device should be checked at least once every 12 months by technical service Every 24 months it is compulsory to have a safety inspection and technical maintenance performed on the device ONLY USE BATTERIES RECOMMENDED BY GIMA USE OF BATTERIES OTHER ...

Page 7: ...ng adapter to the device with the appropriate connector and insert the power cable plug to the power socket To start the treatment press the ON OFF button and check its green back light When the device is switched on the LCD screen activates displaying a bar graph graduated scale and negative pressure value which must be set by the operator The negative pressure is set with keys and which are foun...

Page 8: ... cigarette lighter DC 12V Connect the device s external plug 12V to the ligther plug with the cigarette lighter cable Check the battery power status of the vehicle before the cigarette lighter cable Press the switch to start suction Press the switch to the I position to turn it on WARNING Only use the originally supplied or recommended replacement cigarette ligther cables view the chapter Importan...

Page 9: ...ed Battery operation guaranteed Fixed yellow back light During battery operation Intermediate status Battery not fully charged Battery operation guaranteed start the charge cycle upon activation of the red back light Fixed red back light During battery operation Low battery Start charge phase ATTENTION A long and continuous beep is produced during this signal to warn the operator about the low cha...

Page 10: ...he surgical SUPERVEGA EVO BATTERY ASPIRATOR is intended for use in the electromagnetic environment specifiedbelow The customers or the user of the surgical SUPERVEGA EVO BATTERY ASPIRATOR should assure that it s used in such an environment Emissions test Compliance Electromagnetic environment guidance Irradiated Conducted emissions CISPR11 Group 1 The surgical SUPERVEGA EVO BATTERY ASPIRATOR only ...

Page 11: ...itters and TV transmitters can not be theoretically and accurately foreseen To establish an electro magnetic environment generated by fixed RF transmitters an electro magnetic study of the site should be considered If the field intensity measured in the place where the device will be used surpasses the above mentioned applicable level of conformity the normal functioning of the device should be mo...

Page 12: ...not carried out by technical service RULES FOR RETURNING AND RAPAIRING UNDER NEW EUROPEAN RULES GIMA REQUIRES THE FOLLOWING PROCEDURES TO BECARRIEDOUT TO PROTECT THE INSTRUMENT ANDTHE SAFETY OF ALL WHO COME IN CONTACT WITH IT Before returning an instrument for repair the external surfaces and all accessories MUST be carefully disinfected with a cloth soaked in methylated spirits or hypochlorite ba...

Page 13: ...Gima S p A Via Marconi 1 20060 Gessate MI Italy gima gimaitaly com export gimaitaly com www gimaitaly com PROFESSIONAL MEDICAL PRODUCTS Gima S p A Via Marconi 1 20060 Gessate MI Italy ...

Reviews: