GB
21
After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thouroughly.
Then soak in warm water (temperature shall not exceed 60°C). Wash thouroughly and if necessary use a non-abrasive brush to
remove incrustations. Rinse in running warm water and dry all parts with a soft cloth (non-abrasive).
The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C
(1 bar relative pressure – 15 min) making sure that the jar is positioned upsidedown. Mechanical resistance of the jar is
guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1). Beyond this limit the
physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore recommended.
After sterilization and cooling at environment temperature of the parts make sure that these are not damaged.
The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min).
The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min).
DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER
PERIODICAL MAINTENANCE CHECKS
The
SUPERVEGA EVO BATTERY ASPIRATOR
suction equipment does not need maintenance or lubrication.
It is, however to inspect the unit before each use. With regard to training, given the information contained in the user manual and
since it is easy to understand the said device, it doesn’t appear to be necessary. Unpack the instrument and
always check
integrity
of plastic parts, they might have been damaged during previous use.
Turn switch on and verify that loud noises are not present, these can indicate wrong functioning.
Make sure you do not hear excessively annoying noises that may highlight a malfunction.
The device is protected by a safety fuse (
F 10A L 250V
) situated in the cigarette lighter cable. When replacing, always check the
type and value as indicated.
Internally, the device is protected (see electrical specifications) by two fuses (
T 15A L 125V
) that cannot be reached from the
outside. Therefore, contact the manufacturer to request the assistance of an authorized and qualified technician when they need to
be replaced. If it’s replaced make sure that its replacement is always the same type and value, as indicated.
The device is equipped with a Lithium-Ion Battery, which cannot be reached from the outside. Only refer to
GIMA
technical service
personnel to replace it. The device should be checked at least once every 12 months by technical service. Every 24 months it is
compulsory to have a safety inspection and technical maintenance performed on the device.
ONLY USE BATTERIES RECOMMENDED BY
GIMA.
USE OF BATTERIES OTHER THAN THOSE RECOMMENDED
WILL MAKE THE WARRANTY VOID AND NULL.
ONLY QUALIFIED PERSONNEL ARE AUTHORISED TO REPLACE THE INTERNAL BATTERY. ANY OPERATION
PERFORMED BY NON-TRAINED PERSONNEL MAY CAUSE DANGER (E.G. EXCESSIVE TEMPERATURE)
Fault type
Cause
Solution
1.
Fixed Red Back-Light
Low battery
If it is intended to be used in ambulances, position the
device on the support bracket and leave the battery charge
until the Fixed White Back-Light gives the signal (TAB. I).
2.
No LED is On and the Back-
Light failed to switch on
Locking device
Internal technical problem. Contact technical assistance.
3.
No aspiration
Jar Cap badly screwed down
Unscrewed the cap, then rescrew it correctly
4.
No aspiration
Lid seal not in its seat
Unscrew the cap and insert the seal properly in its seat
5.
The float doesn’t close
If the cap has been washed,
ensure that the float is not
partially detached
Insert the float into it’s place
6.
The float doesn’t close
The float it’s covered by dirty
material
Unscrewed the cap, leave the and put in on autoclave
7.
Low suction
Foam inside the jar
Fill the jar to 1/3 full of ordinary water
8.
No aspiration due to flow
leakage of mucus
Filter blocked
Replace filter
9.
The Vacuum power on the
patient side is either very low
or absent
Protection filter blocked or
damaged
Connection tubes blocked,
kinked or disconnected
Shut-off valve blocked or
damaged
Pump motor damaged
Replace the filter
Replace or reconnect the tubes, check the jar
connections
Empty the jar, or disconnect the tube from the jar and
unblock the shut-off valve. The unit twill only work in
the upright position
Contact the technical service
10.
Noisy
Technical internal problem
Contcat the technical service
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 –
9 - 10
None of the remedies has
achieved the desired results
Contact the seller or
GIMA
After-sales Assistance
Service
Lithium-Ion Battery Charging Cycles:
The Lithium-Ion battery contained inside the device is guaranteed for a number greater
than 300 charging cycles. In the vicinity of 300 charging cycles you can ask for verification of operating status to the manufacturer
or to require replacement of the battery pack in such a way that you always have the component in perfect condition.