Recharging operations:

 to be able to charge the internal battery it is necessary to connect the universal switching adapter  

(mod. UE 60140429SPA1 of FUHUA) to the electric network for approx. 360 minutes with the main switch to position 0.  

The battery’s autonomy when fully charged is approx. 70 minutes with continued operations.  

LIGHT INDICATORS: 

The device is provided with a light indicator (directly on the LCD screen) which allows you to view the operation of the device, 

autonomy of the battery, and the charge phase in progress. 
The light indications, which appear during operation, are indicated in table I. 

The charge cycle starts automatically it is necessary to connect the universal switching adapter (mod. UE 60140429SPA1 of 
FUHUA) to the electric network. The charge occurs only when the motor of the suction unit is off. 

DISPLAY IN CHARGE: this is on with white back-light and it remains on for the entire time. It switches off if there is no external 
voltage. The charge phase lasts until the device is powered and the battery is kept at its maximum charge.  

The large digits display “CHA”

 .  

TAB. I – INDICATOR LIGHTS DURING OPERATIONS 

Back-light signals 

Phase 

Problem / Cause 

Solution 

Fixed white back-light and 

progressive activation of the 
bar-graph 

During the Charge 

phase 

Battery charge in progress 

Wait 

Fixed white back-light and 

fixed activation of the bar-
graph 

During the Charge 

phase 

Charging cycle complete 

Remove the power supply 

Fixed green back-light 

During battery 
operation 

Primary status / Battery 
fully charged 

Battery operation guaranteed 

Fixed yellow back-light 

During battery 
operation 

Intermediate status / 
Battery not fully charged 

Battery operation guaranteed / 
start the charge cycle upon 

activation of the red back-light 

Fixed red back-light 

During battery 
operation 

Low battery 

Start charge phase 

ATTENTION

: A long and 

continuous beep is produced 
during this signal to warn the 

operator about the low charge 
level of the battery. 

Flashing red back-light 

Automatic switch-off of 
the device due to 

discharged battery 

Battery fully discharged 

The flashing back-light switches 
on upon reactivation of the 

device. Immediately start the 
battery charge cycle. 

TAB. II – WRITTEN SIGNALS / BUTTON LED SIGNAL 

Button LED signal / written 
signal 

Function 

Colour 

Position 

ON/OFF button 

Power on 

Green 

Near the front panel key 

Proximity Button 

Switch the ON/OFF key 

Blue 

LED located above the 

“Proximity” key 

CHA 

Displays the battery charge 
status 

Fixed white back-light 
and progressive 

activation of the bar-
graph

 

Central in digits of large 
dimensions 

Small three-digit number 
(000) 

Displays the voltage of the 
device if in charge phase or it 

displays the full scale of the 
bar-graph. 

Fixed white or green 
back-light depending on 

whether external power 
supply or battery and 

progressive activation of 
the bar-graph are used.  

Sideways in small digits 

Bar-graph (arc bar-graph)

 

It displays the trend of the 

negative pressure or of the 
charge status    

Progressive activation of 

the bar-graph (black) 

 

Arc-shaped in the middle of the 

screen 

Large three-digit number 
(000) 

It displays the negative 
pressure set by keys "up" 

and "dw" 

Fixed white or green 
back-light depending on 

whether external power 
supply or battery is used. 

 

Central in digits of large 
dimensions 

Summary of Contents for SUPERVEGA EVO BATTERY ASPIRATOR

Page 1: ...0 Gessate MI Italy gima gimaitaly com export gimaitaly com www gimaitaly com PROFESSIONAL MEDICAL PRODUCTS Gima S p A Via Marconi 1 20060 Gessate MI Italy Made in Italy Instruction manual SUPERVEGA EVO BATTERY ASPIRATOR 230 V 28183 ...

Page 2: ...r immerge the appliance into water Do not place or store the aspirator in places where it may fall or be pulled into the bathtub or washbasin In the event it is accidentally dropped do not attempt to remove the device from the water whilst the plug is still connected disconnect the mains switch remove the plug from the power supply and contact the GIMA technical service department Do not attempt t...

Page 3: ...ormatives IMPORTANT INFORMATION FOR CORRECT DISPOSAL OF THE PRODUCT IN ACCORDANCE WITH EC DIRECTIVE 2012 19 UE RAEE The symbol on the device indicates the separated collection of electric and electronic equipment At the end of life of the device don t dispose it as mixed solid municipal waste but dispose it referring to a specific collection centre located in your area or returring it to the distr...

Page 4: ...t the battery provided with this product shall not be treated as household waste By ensuring these batteries are disposed of correctly you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery The recycling of the materials will help to conserve natural resources At the end of their li...

Page 5: ...ting the said component Conical fitting the number of cycles of sterilization and the number of cleaning cycles is strictly linked to the employment of the said component Therefore after each cleaning cycle it is up to the final user to verify whether the fitting is suitable for reuse The component must be replaced if there are visible signs of decay in the material constituting the said component...

Page 6: ...evice is equipped with a Lithium Ion Battery which cannot be reached from the outside Only refer to GIMA technical service personnel to replace it The device should be checked at least once every 12 months by technical service Every 24 months it is compulsory to have a safety inspection and technical maintenance performed on the device ONLY USE BATTERIES RECOMMENDED BY GIMA USE OF BATTERIES OTHER ...

Page 7: ...ng adapter to the device with the appropriate connector and insert the power cable plug to the power socket To start the treatment press the ON OFF button and check its green back light When the device is switched on the LCD screen activates displaying a bar graph graduated scale and negative pressure value which must be set by the operator The negative pressure is set with keys and which are foun...

Page 8: ... cigarette lighter DC 12V Connect the device s external plug 12V to the ligther plug with the cigarette lighter cable Check the battery power status of the vehicle before the cigarette lighter cable Press the switch to start suction Press the switch to the I position to turn it on WARNING Only use the originally supplied or recommended replacement cigarette ligther cables view the chapter Importan...

Page 9: ...ed Battery operation guaranteed Fixed yellow back light During battery operation Intermediate status Battery not fully charged Battery operation guaranteed start the charge cycle upon activation of the red back light Fixed red back light During battery operation Low battery Start charge phase ATTENTION A long and continuous beep is produced during this signal to warn the operator about the low cha...

Page 10: ...he surgical SUPERVEGA EVO BATTERY ASPIRATOR is intended for use in the electromagnetic environment specifiedbelow The customers or the user of the surgical SUPERVEGA EVO BATTERY ASPIRATOR should assure that it s used in such an environment Emissions test Compliance Electromagnetic environment guidance Irradiated Conducted emissions CISPR11 Group 1 The surgical SUPERVEGA EVO BATTERY ASPIRATOR only ...

Page 11: ...itters and TV transmitters can not be theoretically and accurately foreseen To establish an electro magnetic environment generated by fixed RF transmitters an electro magnetic study of the site should be considered If the field intensity measured in the place where the device will be used surpasses the above mentioned applicable level of conformity the normal functioning of the device should be mo...

Page 12: ...not carried out by technical service RULES FOR RETURNING AND RAPAIRING UNDER NEW EUROPEAN RULES GIMA REQUIRES THE FOLLOWING PROCEDURES TO BECARRIEDOUT TO PROTECT THE INSTRUMENT ANDTHE SAFETY OF ALL WHO COME IN CONTACT WITH IT Before returning an instrument for repair the external surfaces and all accessories MUST be carefully disinfected with a cloth soaked in methylated spirits or hypochlorite ba...

Page 13: ...Gima S p A Via Marconi 1 20060 Gessate MI Italy gima gimaitaly com export gimaitaly com www gimaitaly com PROFESSIONAL MEDICAL PRODUCTS Gima S p A Via Marconi 1 20060 Gessate MI Italy ...

Reviews: