background image

15

ENGLISH

SUPERVEGA BATTERY SUCTION UNIT is a surgical suction pump with the following electrical character-

istics: 14V

 

 

4A with AC/DC (input: 100-240V~ - 50-60Hz – 1.5A) to be used for nasal, oral and tracheal 

aspirations in adults or for body liquids in children (for example mucus, phlegm and blood) equipped with 5 

wheels with braking device. Thanks to these characteristics and its high performance levels, this device is 

particularly suitable for use in hospital wards, in small surgical applications and post-operative treatments. A 

device designed to offer ease-of-transport and almost continuous use thanks to the adoption of an electronic 

system to manage the power supply. Supplied with an audible alarm and visual tell-tale (luminous LED) to 

indicate the battery status. Featuring a body in plastic with high thermal and electrical insulation in compliance 

with recently introduced European safety regulations. Supplied with no. 2 complete sterilisable polycarbonate 

jugs with overflow valve. Features a suction regulator and vacuum gauge located on the front panel.

 GENERAL WARNING

Read instruction manual carefully before use. 

Only highly qualified staff use reserved. 

The instrument must not be disassembled. For a technical service always contact Gima. 

Keep off the reach of children or not capable people without supervision. 

Full containers must be handled with great care during transfer to the disposal areas, following the local 

procedures and regulations.

IMPORTANT SAFETY RULES

1. Check the condition of the unit before each use. The surface of the unit should carefully be inspected for 

visual damage. Check the mains cable and do not connect to power if damage is apparent; 

2. Before connecting the appliance always check that the electric data indicated on the data label and the type 

of plug used, correspond to those of the mains electricity to which it’s to be connected; 

3. Respect the safety regulations indicated for electrical appliances and particularly:

-  Use original components and accessories provided by the manufacturer to guarantee the highest effi-

ciency and safety of the device; 

-  The device can be used only with the bacteriological filter;

-  Never immerge the appliance into water;

-  Position the device on stable and flat surfaces in a way that the air inlets on the back aren’t obstructed;

-  To avoid incidents, do not place the aspirator on unstable surfaces, which may cause it to accidentally fall 

and lead to a malfunction and/or breakage. Should there be signs of damage to the plastic parts, which 

may expose inner parts of the energised device, do not connect the plug to the electrical socket. Do not 

attempt to make the device work before it has been thoroughly checked by qualified personnel and/or the 

GIMA technical service department. 

-  Don’t use in the presence of inflammable substances such as anaesthetic, oxygen or nitrous oxide;

-  Don’t touch the device with wet hands and always prevent the appliance coming into contact with liquids;

-  Don’t leave the appliance connected to the power supply socket when not in use;

-  Don’t pull the power supply cable to disconnect the plug remove the plug from the mains socket correctly;

-  Store and use the device in places protected against the weather and far from any sources of heat. After 

each use, it is recommended to store the device in its own box away from dust and sunlight. 

-  In general, it is inadvisable to use single or multiple adapters and/or extensions. Should their use be 

necessary, you must use ones that are in compliance with safety regulations, however, taking care not to 

exceed the maximum power supply tolerated, which is indicated on the adapters and extensions.

-  Prevent children from using the device without proper supervision;

4. For repairs, exclusively contact technical service and request the use of original spare parts. Failure to 

comply with the above can jeopardise the safety of the device; 

5. Use only for the purpose intended. Don’t use for anything other than the use defined by the manufacturer. 

The manufacturer will not be responsible for damage due to improper use or connection to an electrical 

system not complying with current regulation. 

6. The medical device requires special precautions regarding electromagnetic compatibility and must be in-

stalled and used in accordance with the information provided with the accompanying documents: the SU-

PERVEGA BATTERY SUCTION UNIT device must be installed and used away from mobile and portable 

RF communication devices (mobile phones, transceivers, etc.) that may interference with the said device. 

7. WARNING: Do not change this equipment without the permission of the manufacturer GIMA S.p.A. None 

of electric or mechanical parts has been designed to be repaired by customers or end-users. Don’t open 

Summary of Contents for SUPERVEGA 28193

Page 1: ...EUR SUPERVEGA BATTERY SUR CHARIOT ABSAUGER SUPERVEGA BATTERY AUF TROLLEY ASPIRADOR SUPERVEGA BATTERY SOBRE CARRO M28193 M Rev 1 01 21 PROFESSIONAL MEDICAL PRODUCTS 0476 Gima S p A Via Marconi 1 20060...

Page 2: ...inlets on the back aren t obstructed To avoid incidents do not place the aspirator on unstable surfaces which may cause it to accidentally fall and lead to a malfunction and or breakage Should there...

Page 3: ...in chest fluids The device must not be used for suction of explosive corrosive or easily flammable liquids SUPERVEGA BATTERY SUCTION UNIT is not suitable for MRI Do not introduce the device in MRI env...

Page 4: ...use The filter is not designed to be decontaminated disassembled and or sterilised If the patient s pathology is known and or no risk of indirect contamination exists it is advisable to replace the f...

Page 5: ...rations according to the following procedure Place the overflow valve into its seat in the cover under VACUUM connector Insert floating valve keeping the o ring towards the opening of the cage Place t...

Page 6: ...ttery pay special attention to the polarity of the same component The polarities are indicated directly on the battery If the overfill security system it s activated don t proceede with the liquid asp...

Page 7: ...se must be used in vertical mode to prevent the action of the backflow valve If this protection is triggered turn the device off and disconnect the pipe connected to the vacuum jar indicated with the...

Page 8: ...able to charge the internal battery it is necessary to connect the universal switching adapter to the electric network for approx 240 minutes with the main switch to position 0 TAB I INDICATOR LIGHTS...

Page 9: ...e float valve intervenes signalling the device is too full the suction source must be disconnected within no more than 5minutes RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES This section...

Page 10: ...nt Immunity Test Level indicated by the EN 60601 1 2 Compliance Level Electromagnetic environment guidance Electrostatic discharge ESD EN 61000 4 2 8kV on contact 15kV in air The device doesn t change...

Page 11: ...mined by an elec tro magnetic study of the sitea could be lower than the level of conformity of each frequency interval b It is possible to check for interference in proximity to devices identified by...

Page 12: ...2 12 23 For transmitters with a maximum nominal output power not shown above the recommended separation distance in metres m can be calculated using the equation applicable to the transmitter frequenc...

Page 13: ...rectly you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery The recycling of the...