background image

19

PERIODICAL MAINTENANCE CHECKS

The SUPERVEGA BATTERY SUCTION UNIT suction equipment does not need maintenance or lubrication. It 

is, however, necessary to inspect the unit before each use. With regard to training, given the information con-

tained in the user manual and since it is easy to understand the said device, it doesn’t appear to be necessary. 

Unpack the instrument and always check integrity of plastic parts and AC/DC switching adapter , feeding ca-

ble, they might have been damaged during previous use. Connect cable to electrical network and turn switch 

on. Close the aspirator outlet with your finger and with suction regulator in maximum vacuum position check 

that the vacuum indicators reaches -80kPa (-0.80 bar) minimum (internal battery). Rotate the knob from right 

to left and check the aspiration regulating control. The vacuum indicator should go down -40kPa (-0.40 bar). 

Verify that loud noises are not present, these can indicate wrong functioning. 

Internally, the device is protected (see electrical specifications) by two fuses F1, F2 (T 15A L 125V) that can-

not be reached from the outside. Therefore, contact the manufacturer to request the assistance of an author-

ized and qualified technician when they need to be replaced. If it’s replaced make sure that its replacement is 

always the same type and value, as indicated. 

The device is made up of a lead battery which cannot be accessed from the outside. In order to replace it, 

consult the technical staff authorised by the manufacturer.

 USE ONLY THE RECOMMENDED BATTERIES FROM GIMA. THE USE OF OTHER BATTERIES 

IS NOT RECOMMENDED AND INVOLVES THE CANCELLATION OF WARRANTY

In the event that the service personnel has to replace the internal battery, pay special attention to the polarity 

of the same component. The + / - polarities are indicated directly on the battery. 

If the overfill security system it’s activated, don’t proceede with the liquid aspiration. If the overfill security 

system doesn’t work there are two cases: 

ENGLISH

Fault type

Cause

Solution

1. Red light on

Battery run down

Hook up the power cord to the electricity mains, 

positioning the equipment power switch on 0.

2. No light

Defective AC/DC adapter or technical 

internal problem

Contact the technical service.

3. No aspiration

Jar Cap badly screwed down

Unscrewed the cap, then rescrew it correctly

4. No aspiration

Lid seal not in its seat

Unscrew the cap and insert the seal properly in 

its seat

5. The float doesn’t close

If  the  cap  has  been  washed,  ensure 

that the float is not partially detached

Insert the float into it’s place

6. The float doesn’t close

The float it’s covered by dirty material Unscrewed the cap, leave the and put in on au-

toclave

7. Low suction

Foam inside the jar

Fill the jar to 1/3 full of ordinary water

8. No aspiration due 

to flow leakage of mucus

Filter blocked

Replace filter

9. The Vacuum power 

on the patient side is either 

very low or absent

•  Vacuum regulator set to minimum

•  Protection filter blocked or damaged

•  Connection tubes blocked, kinked or 

disconnected

•  Shut-off valve blocked or damaged

•  Pump motor damaged

• Turn the vacuum regulator clockwise and 

check the value of the vacuum on the gauge

•  Replace the filter

•  Replace or reconnect the tubes, check the jar 

connections

•  Empty the jar, or disconnect the tube from the 

jar and unblock the shut-off valve. The unit twill 

only work in the upright position

•  Contact the technical service

10. Noisy

Technical internal problem

Contact the technical service

Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 

- 8 - 9 - 10

None of the remedies has achieved 

the desired results

Contact the seller or GIMA After-sales Assis-

tance Service

Summary of Contents for SUPERVEGA 28193

Page 1: ...EUR SUPERVEGA BATTERY SUR CHARIOT ABSAUGER SUPERVEGA BATTERY AUF TROLLEY ASPIRADOR SUPERVEGA BATTERY SOBRE CARRO M28193 M Rev 1 01 21 PROFESSIONAL MEDICAL PRODUCTS 0476 Gima S p A Via Marconi 1 20060...

Page 2: ...inlets on the back aren t obstructed To avoid incidents do not place the aspirator on unstable surfaces which may cause it to accidentally fall and lead to a malfunction and or breakage Should there...

Page 3: ...in chest fluids The device must not be used for suction of explosive corrosive or easily flammable liquids SUPERVEGA BATTERY SUCTION UNIT is not suitable for MRI Do not introduce the device in MRI env...

Page 4: ...use The filter is not designed to be decontaminated disassembled and or sterilised If the patient s pathology is known and or no risk of indirect contamination exists it is advisable to replace the f...

Page 5: ...rations according to the following procedure Place the overflow valve into its seat in the cover under VACUUM connector Insert floating valve keeping the o ring towards the opening of the cage Place t...

Page 6: ...ttery pay special attention to the polarity of the same component The polarities are indicated directly on the battery If the overfill security system it s activated don t proceede with the liquid asp...

Page 7: ...se must be used in vertical mode to prevent the action of the backflow valve If this protection is triggered turn the device off and disconnect the pipe connected to the vacuum jar indicated with the...

Page 8: ...able to charge the internal battery it is necessary to connect the universal switching adapter to the electric network for approx 240 minutes with the main switch to position 0 TAB I INDICATOR LIGHTS...

Page 9: ...e float valve intervenes signalling the device is too full the suction source must be disconnected within no more than 5minutes RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES This section...

Page 10: ...nt Immunity Test Level indicated by the EN 60601 1 2 Compliance Level Electromagnetic environment guidance Electrostatic discharge ESD EN 61000 4 2 8kV on contact 15kV in air The device doesn t change...

Page 11: ...mined by an elec tro magnetic study of the sitea could be lower than the level of conformity of each frequency interval b It is possible to check for interference in proximity to devices identified by...

Page 12: ...2 12 23 For transmitters with a maximum nominal output power not shown above the recommended separation distance in metres m can be calculated using the equation applicable to the transmitter frequenc...

Page 13: ...rectly you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery The recycling of the...

Reviews: