21

Filter assembling

Make sure the filter is assembled with the arrows on the side of the patient.

WARNING: The inside of the medical device must be regularly checked for the presence of liquids 

or other visible contamination (secretions). In the presence of liquids or other visible contamination, 

immediately replace the medical device due to the risk of an insufficient vacuum flow rate. 

These products have been designed, tested and manufactured exclusively for single patient use and 

for a period no longer than 24 hours. 

Operation with Internal Battery 

•  Press the switch in position I to turn the device on (the external power pack doesn’t need to be connected)

•  The fully charged battery life is about 60 minutes with continuous operation.

 WARNING: Before using the device, check the battery power status. Before each use proceed with 

charging the battery. To maintain the device in good conditions, recharge the battery every 3 months (when 

not in use).

Recharging operations: 

to be able to charge the internal battery it is necessary to connect the universal 

switching adapter to the electric network for approx. 240 minutes with the main switch to position 0. 

TAB. I – INDICATOR LIGHTS DURING OPERATIONS

When an external power supply (regardless of the state of the battery charger) and when the device is work-

ing (after having turned it on), the LED stays in a FIXED GREEN position.

ENGLISH

LED Signal

Phase

Problem / Cause

Solution

Flashing Green 

Led

During rechanrge

Battery recharge running Wait

Steady Greed Led During rechanrge

Recharging cycle 

complete

Remove power supply

Steady Red Led

During battery 

operation

Flat battery

Start recharging cycle 

WARNING: During this signal, you will hear a 

long, continuous beep (duration of sound 0.8 

sec / sound frequency: every 8.5 sec), which 

notifies  the  user  regarding  the  battery  dis-

charge.

Flashing Red Led

Device automatically 

turns off when the 

battery is flat

Battery completely flat

When the device is restarted the LED will flash 

red: begin the battery recharge cycle immedi-

ately

Steady Orange Led During battery 

operation

Intermediate status

Guaranteed battery function / Recharge when 

the red LED signal comes on.

Summary of Contents for SUPERVEGA 28193

Page 1: ...EUR SUPERVEGA BATTERY SUR CHARIOT ABSAUGER SUPERVEGA BATTERY AUF TROLLEY ASPIRADOR SUPERVEGA BATTERY SOBRE CARRO M28193 M Rev 1 01 21 PROFESSIONAL MEDICAL PRODUCTS 0476 Gima S p A Via Marconi 1 20060...

Page 2: ...inlets on the back aren t obstructed To avoid incidents do not place the aspirator on unstable surfaces which may cause it to accidentally fall and lead to a malfunction and or breakage Should there...

Page 3: ...in chest fluids The device must not be used for suction of explosive corrosive or easily flammable liquids SUPERVEGA BATTERY SUCTION UNIT is not suitable for MRI Do not introduce the device in MRI env...

Page 4: ...use The filter is not designed to be decontaminated disassembled and or sterilised If the patient s pathology is known and or no risk of indirect contamination exists it is advisable to replace the f...

Page 5: ...rations according to the following procedure Place the overflow valve into its seat in the cover under VACUUM connector Insert floating valve keeping the o ring towards the opening of the cage Place t...

Page 6: ...ttery pay special attention to the polarity of the same component The polarities are indicated directly on the battery If the overfill security system it s activated don t proceede with the liquid asp...

Page 7: ...se must be used in vertical mode to prevent the action of the backflow valve If this protection is triggered turn the device off and disconnect the pipe connected to the vacuum jar indicated with the...

Page 8: ...able to charge the internal battery it is necessary to connect the universal switching adapter to the electric network for approx 240 minutes with the main switch to position 0 TAB I INDICATOR LIGHTS...

Page 9: ...e float valve intervenes signalling the device is too full the suction source must be disconnected within no more than 5minutes RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES This section...

Page 10: ...nt Immunity Test Level indicated by the EN 60601 1 2 Compliance Level Electromagnetic environment guidance Electrostatic discharge ESD EN 61000 4 2 8kV on contact 15kV in air The device doesn t change...

Page 11: ...mined by an elec tro magnetic study of the sitea could be lower than the level of conformity of each frequency interval b It is possible to check for interference in proximity to devices identified by...

Page 12: ...2 12 23 For transmitters with a maximum nominal output power not shown above the recommended separation distance in metres m can be calculated using the equation applicable to the transmitter frequenc...

Page 13: ...rectly you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery The recycling of the...

Reviews: