18

Service life of the device: more than 1000 hours of operation (or 3 years) in accordance with the standard 

conditions of testing and operation. Shelf life: maximum 5 years from the date of manufacture. 

WARNING: The medical device is provided without a specific suction probe. If this device must be 

used with a specific suction probe, the end user is responsible for making sure it complies with the 

EN 10079-1 regulation.

CLEANING OF ACCESSORIES

Before using the device, the manufacturer advises you to clean and/or sterilize the accessories.

Washing and / or cleaning the autoclavable jar as to be carried out as follows:

•  Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminat-

ing substances;

•  Disconnect the tank from the device and remove the said container from the support of the device.

•  Separate all the parts of the cover (overflow device, washer).

•  Disconnect all tubes from the jar and the protection filter

•  Wash each part of the container from secretions under cold running water and then clean every single part 

in hot water (temperature not exceeding 60°C)

•  Once again, carefully wash each single part using, if necessary, a non-abrasive brush to remove any de-

posits.

  Rinse with hot running water and dry all parts with a soft cloth (non-abrasive). It is possible to wash with 

commercial disinfectants by carefully following the instructions and dilution values supplied by the manu-

facturer.

  After cleaning, leave the parts to dry in an open, clean environment.

•  Dispose of the aspiration catheter according to that provided by local laws and regulations.

The silicone aspiration tubes and the conical fitting may be carefully washed in hot water (temperature must 

not exceed 60°C). After cleaning, leave the parts to dry in an open, clean environment.

When cleaning is complete, reassemble the container for liquid aspirations according to the following procedure:

•  Place the overflow valve into its seat in the cover (under VACUUM connector)

•  Insert floating valve keeping the o-ring towards the opening of the cage

•  Place the o-ring into its seat around the cover

•  After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leack-

ages or liquid exit

The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization 

stem cycle at 121°C (1 bar relative pressure – 15 min) making sure that the jar is positioned upsidedown. 

Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated 

conditions (EN ISO 10079-1). Beyond this limit the physical-mechanical characteristics of the plastic may 

decrease and replacement of the part is therefore recommended.

After sterilization and cooling at environment temperature of the parts make sure that these are not damaged.

The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pres-

sure – 15 min).

The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pres-

sure – 15 min).

 DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER  

Disposing of disposable bags: 

If the device is fitted with FLOVAC® disposable collection systems (consisting of a rigid, reusable polycar-

bonate container and a single-use polyethylene collection bag), proceed with the disposal of the bag as follows: 

Deactivate the suction source and remove all tubes connected to the container, paying particular attention to 

avoid accidental contamination. Attach the plugs to the “PATIENT” and “TANDEM” connectors by inserting 

them firmly, paying particular attention to avoid accidental contamination. Take the device to the waste collec-

tion area with all the openings correctly sealed, taking into consideration that the product may potentially be 

infected. Discard the product in compliance with the regulations in force at the hospital.

ENGLISH

Summary of Contents for SUPERVEGA 28193

Page 1: ...EUR SUPERVEGA BATTERY SUR CHARIOT ABSAUGER SUPERVEGA BATTERY AUF TROLLEY ASPIRADOR SUPERVEGA BATTERY SOBRE CARRO M28193 M Rev 1 01 21 PROFESSIONAL MEDICAL PRODUCTS 0476 Gima S p A Via Marconi 1 20060...

Page 2: ...inlets on the back aren t obstructed To avoid incidents do not place the aspirator on unstable surfaces which may cause it to accidentally fall and lead to a malfunction and or breakage Should there...

Page 3: ...in chest fluids The device must not be used for suction of explosive corrosive or easily flammable liquids SUPERVEGA BATTERY SUCTION UNIT is not suitable for MRI Do not introduce the device in MRI env...

Page 4: ...use The filter is not designed to be decontaminated disassembled and or sterilised If the patient s pathology is known and or no risk of indirect contamination exists it is advisable to replace the f...

Page 5: ...rations according to the following procedure Place the overflow valve into its seat in the cover under VACUUM connector Insert floating valve keeping the o ring towards the opening of the cage Place t...

Page 6: ...ttery pay special attention to the polarity of the same component The polarities are indicated directly on the battery If the overfill security system it s activated don t proceede with the liquid asp...

Page 7: ...se must be used in vertical mode to prevent the action of the backflow valve If this protection is triggered turn the device off and disconnect the pipe connected to the vacuum jar indicated with the...

Page 8: ...able to charge the internal battery it is necessary to connect the universal switching adapter to the electric network for approx 240 minutes with the main switch to position 0 TAB I INDICATOR LIGHTS...

Page 9: ...e float valve intervenes signalling the device is too full the suction source must be disconnected within no more than 5minutes RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES This section...

Page 10: ...nt Immunity Test Level indicated by the EN 60601 1 2 Compliance Level Electromagnetic environment guidance Electrostatic discharge ESD EN 61000 4 2 8kV on contact 15kV in air The device doesn t change...

Page 11: ...mined by an elec tro magnetic study of the sitea could be lower than the level of conformity of each frequency interval b It is possible to check for interference in proximity to devices identified by...

Page 12: ...2 12 23 For transmitters with a maximum nominal output power not shown above the recommended separation distance in metres m can be calculated using the equation applicable to the transmitter frequenc...

Page 13: ...rectly you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery The recycling of the...

Reviews: