electromagnetIc compatIBIlIty InformatIon
table 1
For all ME EQUIPMENt and ME SYStEMS
Guidance and manufacture’s declaration - electromagnetic emissions
the KD-595 is intended for use in the electromagnetic environment specified
below. the customer or the user of the KD-595 should assure that it is used
in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
rf emissions CISpr
11
group 1
the KD-595 uses rf energy only for its internal
function. therefore, its rf emissions are very low
and are not likely to cause any interference in
nearby electronic equipment
rf emissions CISpr
11
Class b
the KD-595 is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes
Harmonic emissions
IEC 61000-3-2
not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
not applicable
table 2
For all ME EQUIPMENt and ME SYStEMS
Guidance and manufacturer’s declaration - electromagnetic immunity
the KD-595 is intended for use in the electromagnetic environment specified
below. the customer or the user of the KD-595 should assure that it is used
in such an environment.
IMMUNItY test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV
contact ±
8 kV air
floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%
power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 a/m
3 a/m
power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment