33
ENGLISH
For ME EQUIPMENt and ME SYStEMS that are not LIFE-SUPPOrtING
Guidance and manufacturer’s declaration - electromagnetic immunity
the KD-595 is intended for use in the electromagnetic environment specified
below. the customer or the user of the KD-595 should assure that it is used
in such an environment.
IMMUNItY test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
radiated rf IEC
61000-4-3
3 V/m 80
mHz
to 2.5 gHz
3 V/m
portable and mobile rf
communications equipment should
be used no closer to any part of
the KD-595, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
recommended separation
distance:
d
= 1.2 √
P
80 mHz to 800
mHz
d
= 2.3 √
P
800 mHz to 2,5
gHz
Where
P
is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
d
is
the recommended separation
distance in meters (m).
field strengths from fixed rf
transmitters, as determined by
an electromagnetic site
survey,
a
should be less than
the compliance level in each
frequency range.
b
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
Note 1:
at 80 mHz and 800 mHz, the higher frequency range applies.
Note 2:
these guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.