If the current status is alarm silence, a long or short press of the alarm silence key can de-activate the
alarm silence function.
4.11 Data Uploading
When connected to a computer via USB, the device enters into data uploading mode, as shown in figure 2.25.
Figure 4.25 Data uploading screen
In data uploading mode, the device will automatically stop SpO
2
measurement, NIBP measurement, Hemo-
stat, pressure verification and air leakage checks etc., and all key operations will be disabled except power
key.
Chapter 5
ALARMS
5.1 Alarm Priority
Low Priority:
NIBP over range
Temp over range
ECG Lead off
SpO
2
Probe off
Temp Probe off
NIBP error message
SpO
2
error message
Medium Priority:
HR/PR over range
High Priority:
ECG Arrest
Ventricular Run
Ventricular fibrillation
SpO
2
over limit
SYS over limit
DIA over limit
MAP over limit
Temp over limit
HR/PR over limit
Can not detect SpO
2
Can not detect HR/PR