10
Description to AC, DC Power indicator:
11 MAP/Time:
display mean arterial pressure at the end of a successful measurement and end time (in Ma-
nual) or STAT mode) or counting down time (in Auto mode) alternatively. Cuff pressure is displayed during
BP measurement or the hemostat function is in use.
There are 2 situations of display when the NIBP measurement finishes:
1) When NIBP measurement mode is set to “Manual” or “STAT”, the mean arterial pressure or measuring time
will be displayed alternately, the time format is “hh:mm”.
2) When NIBP measurement mode is set to “AUTO”, the counting down time will be displayed, the time format
is “mm:ss”. If the counting down time is over 1 hour, then it displays “hh:mm”.
Note:
BP value can be displayed in two units, “×××” mmHg” or “××.×” kPa, refer to section “4.9.3 NIBP Setup”
to set the unit of BP value. The conversion between two units is: 1kPa=7.5 mmHg, 1mmHg=0.133kPa.
12
NIBP Setup key:
A shortcut key to change the NIBP measuring mode and cycle time for Auto mode.
13
ECG lead selection key:
A short press of this key to shift ECG lead, holding this key and pressing
NIBP setup key (12) will lock or un-lock the key operation.
14
Patient category indicator:
for adult;
for pediatric;
for neonate.
15
Alarm silence key.
16 Print:
The built-in printer is optional. If installed, press this key to print the current measured data.
17
: Alarm silence indicator.
When it is on, it indicates that the alarm is silenced.
18 NIBP:
press to start/cancel NIBP measurement.
19 :
Bar-graph of pulse intensity.
20
Up:
shifts cursor forward/upward.
21
OK:
In setting menu, press it to confirm selection or modification; On history record screen, long pressing
this key to open up delete data dialogue box; On monitoring screen, short press to freeze/unfreeze ECG
waveform.
22
Down:
shifts cursor backward/downward.
23
Display View key:
short press to scroll through LCD display views or return to the upper level screen;
long press to enter into root setting menu display screen.
24 TEMP:
TEMP probe connector.
25 NIBP:
NIBP hose connector.
26 SpO
2
:
SpO
2
sensor connector.
Note:
A long press would mean press and hold for 2 seconds.
AC Power indicator DC Power indicator
Descriptions
Status
ON
ON
This device is on and using AC power supply.
OFF
ON
The device is on and using built-in battery.
ON
OFF
The device is off and battery is being charged
while the AC power is connected.