45
3. Visual alarm: The visual alarm includes the LED indicator located on the upper front panel of the Gima Vital
Signs Monitor, the numeric readings flashing, and the alarm message displayed on the bottom of the LCD
screen. Alarm indicator frequency and color see below:
Alarm LED indicator:
High-priority: red light flashing with 2Hz frequency and 50% duty ratio
Low priority: Yellow light flashing with 2Hz frequency and 50% duty ratio.
Medium priority: Yellow light on.
No alarm: Green light on.
Numeric reading alarm:
the reading value flashing reversed color display.
4. Alarm reset and silence: see Section 4.10.
6.10.6 Additional description for power supply, network and display
1. Power supply: main power supply: AC 100V~240V, 50Hz/60Hz
Internal power supply: 11.1VDC.
2. Input power: <50VA.
3. The minimum working time when operating with all accessories by internal power supply: 270min.
4. Network connection: Ethernet network.
5. Display panel: color TFT LCD.
6. Working modes: Demo mode and Real-time mode.
6.11 Guidance and manufacturer’s declaration-Electromagnetic compatibility
Table 1
Guidance and manufacturer’s declaration-electromagnetic emission for all EQUIPMENT AND SYSTEM
Vital Signs Monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment or system should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
Vital Signs Monitor uses RF energy only for its in-
ternal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class A
Vital Signs Monitor is suitable for use in all esta-
blishments other than domestic and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC61000-3-3
Complies