38
off the pulse beep sound, the factory default is set at"2”. The tone of pulse beat beep changes when the
measured SpO
2
changes i.e. the higher the SpO
2
value is, the higher the tone of the pulse beep (which be-
comes sharper); and the lower the SpO
2
value is, the lower the tone of the pulse beep.
•
Print Mode:
sets the printing time for real-time printing mode with the options of “Continue”, “10s”, “20s”,
“30s” and “60s” for optional. “Continue” means that the device will not stop printing the real-time plethysmo-
gram and ECG waveform until the user change the display screen or press the print key again.
XXs:
prints the real-time plethysmogram and ECG waveform lasting for XX seconds.
Note:
the gray background means this item is nonadjustable.
4.9.11 Reset to Factory Default Settings
Figure 4.23 Default setting
4.9.12 About
This displays the software version and serial number, as shown in figure 4.24, refer to the Gima Vital Signs
Monitor for detailed information.
Figure 4.24 About
4.10 Alarm Settings
Press the alarm silence
key to set the alarm sound status. There are 3 options in total:
• Alarm sound is enabled, which is the default status.
• Short term alarm silence (120 seconds): a short press of the alarm silence key will display the red icon
,
on the lower screen and the message “silence count-down time 120”. At this time, the alarm silence indica-
tor on the left side of the alarm silence key will be lit. The device will mute the alarm sound temporarily for 2
minutes, but keep the visual alarm flashing. When the time (120s) is up, the alarm silence will be de-activated
automatically, the red icon
will disappear as well, and the alarm silence indicator will be dark.
• Long term alarm silence: long press the alarm silence key will be display the red icon
on the lower screen
and the alarm silence indicator on the left side of the alarm silence key will be lit. The device will mute the
alarm sound on an ongoing basis but keep the visual alarm flashing untill a new type of alarm event is de-
tected. The alarm silence status will then be terminated automatically and the alarm sound will resume, the
red icon
will disappear as well, and the alarm silence indicator will be dark.