Guidance and manufacturer’s declaration – electromagnetic immunity

The device is intended for use in the electromagnetic environments listed below, and 
should only be used in such environments:

38

ENGLISH

IMMUNITY 

TEST

IEC 60601 

TEST LEVEL

COMPLIANCE 

LEVEL

ELECTROMAGNETIC ENVIRON-

MENT  GUIDANCE

Conducted RF
IEC 61000-4-6

3V rms
At 0.15-80 MHz

6V rms
At ISM & Radio 
Amateur Freq.

3V rms
At 0.15-80 MHz

6V rms
At ISM & Radio 
Amateur Freq.

Portable and mobile RF communi-
cations equipment should be used 
no closer to any part of the device, 
including cables, than the recom-
mended separation distance calculat-
ed from the equation applicable to the 
frequency of the transmitter.

Radiated RF
IEC 61000-4-3
(Proximity 
fields from RF 
wireless com-
munications 
equipment
IEC 61000-4-3)

110 V/m at 
80-2700 MHz 
AM Modulation 
And 9-28V/m 
at 385-6000MHz,
Pulse Mode and 
other Modulation. 
The system shall 
be tested as 
specified in 
IEC60601-1-2 
Table 9 for proximity 
fields from RF 
wireless communi-
cations equipment 
using the test 
methods specified 
in IEC 61000-4-3

10 V/m at 
80-2700 MHz 
AM Modulation 
And 9-28V/m 
at 385-6000MHz,
Pulse Mode and 
other Modulation. 
The system shall 
be tested as 
specified in 
IEC60601-1-2 
Table 9 for proximity 
fields from RF 
wireless communi-
cations equipment 
using the test 
methods specified 
in IEC 61000-4-3

Recommended separation distance

Considering to reduce the minimum 
separation distance, based on RISK 
MANAGEMENT, and using higher IM-
MUNITY TEST LEVELS that are appro-
priate for the reduced minimum sepa-
ration distance. Minimum separation 
distances for higher IMMUNITY TEST 
LEVELS shall be calculated using the 
following equation:

E = 6/d  

where 

P

 is the maximum power in W, 

d

 

is the minimum separation distance in 
m, and 

E

 is the IMMUNITY TEST LEV-

ELS in V/m.
Field strengths from fixed RF transmit-
ters, as determined by an electromag-
netic site survey,a should be less than 
the compliance level in each frequency 
range.

b

Interference may occur 
in the vicinity of equip-
ment marked with the 
following symbol: 

NOTA 1 

At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2

 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.

a.

 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones 

and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be 
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF 
transmitters, an electromagnetic site survey should be considered. If the measured field strength in 
the location in which the device is used exceeds the applicable RF compliance level above, the device 
should be observed to verify normal operation. If abnormal performance is observed, additional meas-
ures may be necessary, such as reorienting or relocating the device.

b. 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

√P

Summary of Contents for 32924n

Page 1: ...NSIOM TRE BRACHIAL DE LA PRESSION ART RIELLE MONITOR AUTOM TICO DE BRAZO DE PRESI N SANGU NEA M32924 M Rev 5 02 22 PROFESSIONAL MEDICAL PRODUCTS 0476 Gima S p A Via Marconi 1 20060 Gessate MI Italy gi...

Page 2: ...ded by your doctor You are not to use the information contained herein or this product for diagnosing or treat ing a health problem or prescribing any medication If you have or suspect that you have a...

Page 3: ...clinic is often asso ciated with a phenomenon called White Coat Hypertension where the patient becomes nervous or anxious thus raising his blood pressure There are also numerous other factors that mig...

Page 4: ...blood pres sure as well as your heart rate It s recommended for use by people over the age of 18 and not to be used on infant or children The device is designed for home use and not suitable for clin...

Page 5: ...ofessional before using the device This product is not suitable for people with arrhythmias and pregnant women Blood pressure measurements taken with this device are equivalent to those obtained by a...

Page 6: ...mponents Caution Substitution of a component different from that supplied might result in measurement error 25 ENGLISH 6V 1A AC Adapter Excluded MEMORY BUTTON AA 1 5V ALKALINE BATTERY X 4 START STOP B...

Page 7: ...ned Pulse Symbol Once pulse is detected the symbol flashes with each pulse beat Our suggestion Please do not talk or move during meas urements Irregular Heartbeat Detector This symbol appears for 1 mi...

Page 8: ...gularity is detected during measurement often we recommend you seek medical advice As a safeguard we recommend that if you have arrhythmias such as atrial or ventricular premature beats and atrial fib...

Page 9: ...59 90 99 Red Provide advice about lifestyle modification Hypertension and confirm within two months High Normal 130 139 85 89 Orange Provide advice about lifestyle modification and recheck in one year...

Page 10: ...ith new ones and do not mix new and old batteries Do not mix alkaline standard carbon zinc or rechargeable cadmium batteries either It may shorten the battery life or cause the device to malfunction R...

Page 11: ...rachial artery approximately 1 inch 2 3 cm above the elbow on the inside of your left arm to determine where your strongest pulse is Slide the end of arm cuff furthest from the tube through the metal...

Page 12: ...our arm is wounded or injured MEASUREMENT PROCEDURE Switch on the monitor A Press button to switch on the monitor B All segments appear on the screen Setting year date and time A Press button YEAR fla...

Page 13: ...ector symbol with date and time for 1 minute E Without any operation for 1 minute device automatically shuts off MEMORY FUNCTION Storing data After each measurement the systolic and diastolic pressure...

Page 14: ...ng chemicals to clean the device Use only a dry cloth to wipe the cuff Do not attempt to disassemble or change any parts of the monitor including arm cuff due to substitution of a component different...

Page 15: ...ol appears when blood pressure value displayed is excessively low or high Cuff has been placed incorrectly Wrap the cuff properly so that it is positioned correctly Did you talk or move during measure...

Page 16: ...Total for 3 Users Unit Dimensions 5 51 x 4 09 x 1 97 inch L x W x H 140 x 104 x 50 mm L x W x H Unit Weight 8 73 0 35 oz 247 5 10 g Cuff and Batteries Excluded Cuff Size 23 33 cm 9 13 inch Storage Tr...

Page 17: ...caused by electromagnetic interference between elec trical and electronic equipments do not use the device near a mobile phone or microwave oven At least keep a maximum output power of 2 W yields and...

Page 18: ...8 kV contact discharge 15 kV air discharge In the case of air discharge testing the climatic conditions shall be within the following ranges Ambient Temperature 15 C 35 C Relative Humidity 30 60 Power...

Page 19: ...distance based on RISK MANAGEMENT and using higher IM MUNITY TEST LEVELS that are appro priate for the reduced minimum sepa ration distance Minimum separation distances for higher IMMUNITY TEST LEVELS...

Page 20: ...18 Hz 28 870 930 1720 Pulse modulation 217 Hz 28 1845 1970 2450 Pulse modulation 217 Hz 28 5240 Pulse modulation 217 Hz 9 5500 5785 NOTE If necessary to achieve the IMMUNITY TEST LEVEL the distance b...