APPENDIX

Guidance and manufacturer’s declaration – electromagnetic emissions 

The device is intended for use in the electromagnetic environments listed below, and 
should only be used in such environments:

Guidance and manufacturer’s declaration – electromagnetic immunity

The device is intended for use in the electromagnetic environments listed below, and 
should only be used in such environments:

Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions CISPR 11

Group 1

RF energy is used only to maintain device’s oper-
ation. Therefore, its RF emissions are so low that 
it’s not likely to cause any interference in nearby 
electronic equipment. 

RF emissions CISPR 11

Class B

The device is suitable for use in all establish-
ments, including domestic establishments, and 
those directly connected to the public low-voltage 
power supply network that supplies buildings 
used for domestic purposes.

Harmonic emissions 
IEC 61000-3-2

Class A

Voltage fluctuations/ flicker 
emissions IEC 61000-3-3

Complies

IMMUNITY TEST IEC 60601 

TEST LEVEL

COMPLIANCE 

LEVEL

ELECTROMAGNETIC ENVIRON-

MENT –GUIDANCE

Electrostatic
discharge (ESD)
IEC 61000-4-2

± 8 kV contact 
discharge

± 15 kV air 
discharge

± 8 kV contact 
discharge

± 15 kV air 
discharge

In the case of air discharge testing, 
the climatic conditions shall be 
within the following ranges:
Ambient Temperature:15°C~35°C, 
Relative Humidity: 30%~60%.

Power frequency
(50 or 60 Hz)
magnetic field
IEC 61000-4-8

30 A/m
50 or 60 Hz

30 A/m
50 or 60 Hz

Power frequency magnetic fields 
should be at levels characteristic of 
a typical location in a typical com-
mercial or hospital environment. 

Electrical fast
transient/burst
IEC 61000-4-4

± 2 kV for power 
supply lines

± 1 kV for input/out-
put lines

± 2 kV for power 
supply lines

± 1 kV for input/
output lines

Mains power quality should be that 
of a typical commercial or hospital 
environment.

Surge
IEC 61000-4-5

AC Power port
±1 KV Line to Line

AC Power port
±1 KV Line to Line

Mains power quality should be that 
of a typical commercial or hospital 
environment.

Interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11

0% 

U

T

; 0.5 cycle

At 0°,45°, 90°, 
135°, 180°, 225°, 
270°and 315°.
0% 

U

T

; 1 cycles

70% 

U

T

25/30 cycles
0% 

U

T

250/300 cycle

0% 

U

T

; 0.5 cycle

At 0°,45°, 90°, 135°, 
180°, 225°, 270° 
and 315°.
0% 

U

T

; 1 cycles

70% 

U

T

25/30 cycles
0% 

U

T

250/300 cycle

Mains power quality should be that 
of a typical commercial or hospital 
environment. If the user of the de-
vice requires continued operation 
during power mains interruptions, 
it is recommended that the device 
be powered from an uninterruptible 
power supply or a battery.

Summary of Contents for 32924n

Page 1: ...NSIOM TRE BRACHIAL DE LA PRESSION ART RIELLE MONITOR AUTOM TICO DE BRAZO DE PRESI N SANGU NEA M32924 M Rev 5 02 22 PROFESSIONAL MEDICAL PRODUCTS 0476 Gima S p A Via Marconi 1 20060 Gessate MI Italy gi...

Page 2: ...ded by your doctor You are not to use the information contained herein or this product for diagnosing or treat ing a health problem or prescribing any medication If you have or suspect that you have a...

Page 3: ...clinic is often asso ciated with a phenomenon called White Coat Hypertension where the patient becomes nervous or anxious thus raising his blood pressure There are also numerous other factors that mig...

Page 4: ...blood pres sure as well as your heart rate It s recommended for use by people over the age of 18 and not to be used on infant or children The device is designed for home use and not suitable for clin...

Page 5: ...ofessional before using the device This product is not suitable for people with arrhythmias and pregnant women Blood pressure measurements taken with this device are equivalent to those obtained by a...

Page 6: ...mponents Caution Substitution of a component different from that supplied might result in measurement error 25 ENGLISH 6V 1A AC Adapter Excluded MEMORY BUTTON AA 1 5V ALKALINE BATTERY X 4 START STOP B...

Page 7: ...ned Pulse Symbol Once pulse is detected the symbol flashes with each pulse beat Our suggestion Please do not talk or move during meas urements Irregular Heartbeat Detector This symbol appears for 1 mi...

Page 8: ...gularity is detected during measurement often we recommend you seek medical advice As a safeguard we recommend that if you have arrhythmias such as atrial or ventricular premature beats and atrial fib...

Page 9: ...59 90 99 Red Provide advice about lifestyle modification Hypertension and confirm within two months High Normal 130 139 85 89 Orange Provide advice about lifestyle modification and recheck in one year...

Page 10: ...ith new ones and do not mix new and old batteries Do not mix alkaline standard carbon zinc or rechargeable cadmium batteries either It may shorten the battery life or cause the device to malfunction R...

Page 11: ...rachial artery approximately 1 inch 2 3 cm above the elbow on the inside of your left arm to determine where your strongest pulse is Slide the end of arm cuff furthest from the tube through the metal...

Page 12: ...our arm is wounded or injured MEASUREMENT PROCEDURE Switch on the monitor A Press button to switch on the monitor B All segments appear on the screen Setting year date and time A Press button YEAR fla...

Page 13: ...ector symbol with date and time for 1 minute E Without any operation for 1 minute device automatically shuts off MEMORY FUNCTION Storing data After each measurement the systolic and diastolic pressure...

Page 14: ...ng chemicals to clean the device Use only a dry cloth to wipe the cuff Do not attempt to disassemble or change any parts of the monitor including arm cuff due to substitution of a component different...

Page 15: ...ol appears when blood pressure value displayed is excessively low or high Cuff has been placed incorrectly Wrap the cuff properly so that it is positioned correctly Did you talk or move during measure...

Page 16: ...Total for 3 Users Unit Dimensions 5 51 x 4 09 x 1 97 inch L x W x H 140 x 104 x 50 mm L x W x H Unit Weight 8 73 0 35 oz 247 5 10 g Cuff and Batteries Excluded Cuff Size 23 33 cm 9 13 inch Storage Tr...

Page 17: ...caused by electromagnetic interference between elec trical and electronic equipments do not use the device near a mobile phone or microwave oven At least keep a maximum output power of 2 W yields and...

Page 18: ...8 kV contact discharge 15 kV air discharge In the case of air discharge testing the climatic conditions shall be within the following ranges Ambient Temperature 15 C 35 C Relative Humidity 30 60 Power...

Page 19: ...distance based on RISK MANAGEMENT and using higher IM MUNITY TEST LEVELS that are appro priate for the reduced minimum sepa ration distance Minimum separation distances for higher IMMUNITY TEST LEVELS...

Page 20: ...18 Hz 28 870 930 1720 Pulse modulation 217 Hz 28 1845 1970 2450 Pulse modulation 217 Hz 28 5240 Pulse modulation 217 Hz 9 5500 5785 NOTE If necessary to achieve the IMMUNITY TEST LEVEL the distance b...

Reviews: