Test frequency (MHz)
Modulation
IMMUNITY TEST LEVEL (V/m)
385
Pulse modulation 18 Hz
27
450
FM ± 5 kHz deviation 1kHz sine
28
710
Pulse modulation 217 Hz
9
745
780
810
Pulse modulation 18 Hz
28
870
930
1720
Pulse modulation 217 Hz
28
1845
1970
2450
Pulse modulation 217 Hz
28
5240
Pulse modulation 217 Hz
9
5500
5785
NOTE:
If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m, the 1 m test distance is
permitted by IEC 61000-4-3.
a). The carrier shall be modulated using a 50% duty cycle square wave signal.
b). AS an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it
does not represent actual modulation, it would be worst case.
Test specifications for enclosure port immunity to RF wireless communications equip-