background image

Risk Category Indicator

This device is equipped with Risk Category Indicator which classifies your blood pressure 
measurements into six stages (Optimal to Severe hypertension) as shown in below chart:

Stages of Blood  

Systolic  Diastolic  Color 

Recommendations by SIGN n. 49:

Pressure Levels 

(mmHg)  (mmHg)  

 

Hypertension in older people

Grade 3  Severe 

≥180 

≥110 

Red 

Confirm immediately and repeat BP in one

 

Hypertension 

 

 

 

day and again within one week depending

 

 

 

 

 

on clinical situation. 

Grade 2  Moderate 

160 ~ 179  100 ~ 109 

Red 

Serial blood pressures repeated within 

 

Hypertension 

 

 

 

one month.

Grade 1  Mild 

140 ~ 159  90 ~ 99 

Red 

Provide advice about lifestyle modification

 

Hypertension 

 

 

 

and confirm within two months. 

High-Normal 

130 ~ 139  85 ~ 89  Orange  Provide advice about lifestyle modification

 

 

 

 

 

and recheck in one year.

Normal 

 

120 ~ 129  80 ~ 84 

Yellow  Recheck in 2 - 5 years.

 

 

 

 

 

(patients aged > 75 years offered annual 

Optimal   

< 120 

< 80 

Green 

health check).

 

Source: WHO 2003 

After each measurement is completed, LCD display will show your position automatically 
on the six segments of the bar indicator which corresponds to Risk Category Indicator.

*Note!

When a person’s systolic and diastolic pressures fall into different categories, the higher 
category should apply.
e.g. systolic rate 181 & diastolic rate 99  

 

Red category (Severe Hypertension)

e.g. systolic rate 110 & diastolic rate 95  

 

Red category (Mild Hypertension)

28

 

ENGLISH

Summary of Contents for 32924n

Page 1: ...NSIOM TRE BRACHIAL DE LA PRESSION ART RIELLE MONITOR AUTOM TICO DE BRAZO DE PRESI N SANGU NEA M32924 M Rev 5 02 22 PROFESSIONAL MEDICAL PRODUCTS 0476 Gima S p A Via Marconi 1 20060 Gessate MI Italy gi...

Page 2: ...ded by your doctor You are not to use the information contained herein or this product for diagnosing or treat ing a health problem or prescribing any medication If you have or suspect that you have a...

Page 3: ...clinic is often asso ciated with a phenomenon called White Coat Hypertension where the patient becomes nervous or anxious thus raising his blood pressure There are also numerous other factors that mig...

Page 4: ...blood pres sure as well as your heart rate It s recommended for use by people over the age of 18 and not to be used on infant or children The device is designed for home use and not suitable for clin...

Page 5: ...ofessional before using the device This product is not suitable for people with arrhythmias and pregnant women Blood pressure measurements taken with this device are equivalent to those obtained by a...

Page 6: ...mponents Caution Substitution of a component different from that supplied might result in measurement error 25 ENGLISH 6V 1A AC Adapter Excluded MEMORY BUTTON AA 1 5V ALKALINE BATTERY X 4 START STOP B...

Page 7: ...ned Pulse Symbol Once pulse is detected the symbol flashes with each pulse beat Our suggestion Please do not talk or move during meas urements Irregular Heartbeat Detector This symbol appears for 1 mi...

Page 8: ...gularity is detected during measurement often we recommend you seek medical advice As a safeguard we recommend that if you have arrhythmias such as atrial or ventricular premature beats and atrial fib...

Page 9: ...59 90 99 Red Provide advice about lifestyle modification Hypertension and confirm within two months High Normal 130 139 85 89 Orange Provide advice about lifestyle modification and recheck in one year...

Page 10: ...ith new ones and do not mix new and old batteries Do not mix alkaline standard carbon zinc or rechargeable cadmium batteries either It may shorten the battery life or cause the device to malfunction R...

Page 11: ...rachial artery approximately 1 inch 2 3 cm above the elbow on the inside of your left arm to determine where your strongest pulse is Slide the end of arm cuff furthest from the tube through the metal...

Page 12: ...our arm is wounded or injured MEASUREMENT PROCEDURE Switch on the monitor A Press button to switch on the monitor B All segments appear on the screen Setting year date and time A Press button YEAR fla...

Page 13: ...ector symbol with date and time for 1 minute E Without any operation for 1 minute device automatically shuts off MEMORY FUNCTION Storing data After each measurement the systolic and diastolic pressure...

Page 14: ...ng chemicals to clean the device Use only a dry cloth to wipe the cuff Do not attempt to disassemble or change any parts of the monitor including arm cuff due to substitution of a component different...

Page 15: ...ol appears when blood pressure value displayed is excessively low or high Cuff has been placed incorrectly Wrap the cuff properly so that it is positioned correctly Did you talk or move during measure...

Page 16: ...Total for 3 Users Unit Dimensions 5 51 x 4 09 x 1 97 inch L x W x H 140 x 104 x 50 mm L x W x H Unit Weight 8 73 0 35 oz 247 5 10 g Cuff and Batteries Excluded Cuff Size 23 33 cm 9 13 inch Storage Tr...

Page 17: ...caused by electromagnetic interference between elec trical and electronic equipments do not use the device near a mobile phone or microwave oven At least keep a maximum output power of 2 W yields and...

Page 18: ...8 kV contact discharge 15 kV air discharge In the case of air discharge testing the climatic conditions shall be within the following ranges Ambient Temperature 15 C 35 C Relative Humidity 30 60 Power...

Page 19: ...distance based on RISK MANAGEMENT and using higher IM MUNITY TEST LEVELS that are appro priate for the reduced minimum sepa ration distance Minimum separation distances for higher IMMUNITY TEST LEVELS...

Page 20: ...18 Hz 28 870 930 1720 Pulse modulation 217 Hz 28 1845 1970 2450 Pulse modulation 217 Hz 28 5240 Pulse modulation 217 Hz 9 5500 5785 NOTE If necessary to achieve the IMMUNITY TEST LEVEL the distance b...

Reviews: